User manual
Table Of Contents
- 1. BASIC INFORMATION
 - 2. TRANSPORT AND STORAGE
 - 3. UNPACKING, INSTALLATION AND COMMISSIONING
 - 4. GETTING STARTED
 - 5. BALANCE LEVELING
 - 6. MENU STRUCTURE
 - 7. WEIGHING
 - 8. BASICS ON TERMINAL AND SOFTWARE OEPRATION
 - 9. BALANCE SOFTWARE
 - 10. BALANCE ADJUSTMENT
 - 11. DISPLAY CALIBRATION
 - 12. BALANCE USER SELECTION
 - 13. PERIPHERALS
 - 14. ABOUT
 - 15. COMMUNICATION
 - 16. UPDATE
 - 17. LEVELING
 - 18. BALANCE ADAPTATION – USER SETTINGS
- 18.1. Adding a new operator
 - 18.2. Deleting a operator
 - 18.3. Selecting a operator
 - 18.4. Operator settings
 - 18.5. Quick access to determining working modes parameters
 - 18.6. Parameter of working mode WEIGHING – quick access keys
 - 18.7. Parameter of working mode WEIGHING – information
 - 18.8. Parts Counting
 - 18.9. Checkweighing
 - 18.10. Dosing
 - 18.11. Percent setup
 - 18.12. Animal weighing
 - 18.13. Density of solids, liquid and air
 - 18.14. Formulations
 - 18.15. Statistics
 - 18.16. Mass Comparison
 - 18.17. Differential weighing
 
 - 19. ACCESS TO BALANCE PARAMETERS
 - 20. TECHNICAL DATA
 - APPENDIX A – cooperation with EPSON printer
 - APPENDIX B – cooperation with KAFKA printer
 - APPENDIX C – cooperation with label printer CITIZEN
 - APPENDIX D – databases
 

- 62 - 
18.8.3. ACAI function – Automatic Accuracy Correction 
The ACAI function is to automatically correct mass of a single part during 
parts counting process. it is aimed to increase the accuracy of counting 
process – mass of a single part is calculated from a higher quantity of 
samples, ensuring higher accuracy.  
First, enable the ACAI function. 
PROCEDURE: 
o In an optional moment press 
grey coloured workspace area 
o Which opens a screen on 
balance internal menu 
o Scroll the menu to option ACAI 
o Press ACAI button 
o In an opened message box select option YES 
Principles of using the ACAI 
function: 
1.  Number of parts (on adding) on 
balance’s weighing pan has to 
be greater than has been 
previously 
2.  number of parts (on adding) on 
balance’s weighing pan must be 
less than twice the amount of 
which was visible on the display 
before adding parts 
3.  current quantity of parts must be within the 
 0         
value, 
4.  measurement result has to be stabilized. 
Enabled ACAI function is indicated 
by a black pictogram displayed 
next to the number counted of 
parts. If information item “Standard 
mass” is enabled in the internal 
mode menu (access by pressing 
the workspace area), then in 
meeting the requirements 1-4 the 
displayed mass of a single part is 
subject to automatic accuracy 
correction. 
This value can be saved in the “Database of Standards” using a 
corresponding quick access key. 