User manual
Table Of Contents
- 1. BASIC INFORMATION
 - 2. TRANSPORT AND STORAGE
 - 3. UNPACKING, INSTALLATION AND COMMISSIONING
 - 4. GETTING STARTED
 - 5. BALANCE LEVELING
 - 6. MENU STRUCTURE
 - 7. WEIGHING
 - 8. BASICS ON TERMINAL AND SOFTWARE OEPRATION
 - 9. BALANCE SOFTWARE
 - 10. BALANCE ADJUSTMENT
 - 11. DISPLAY CALIBRATION
 - 12. BALANCE USER SELECTION
 - 13. PERIPHERALS
 - 14. ABOUT
 - 15. COMMUNICATION
 - 16. UPDATE
 - 17. LEVELING
 - 18. BALANCE ADAPTATION – USER SETTINGS
- 18.1. Adding a new operator
 - 18.2. Deleting a operator
 - 18.3. Selecting a operator
 - 18.4. Operator settings
 - 18.5. Quick access to determining working modes parameters
 - 18.6. Parameter of working mode WEIGHING – quick access keys
 - 18.7. Parameter of working mode WEIGHING – information
 - 18.8. Parts Counting
 - 18.9. Checkweighing
 - 18.10. Dosing
 - 18.11. Percent setup
 - 18.12. Animal weighing
 - 18.13. Density of solids, liquid and air
 - 18.14. Formulations
 - 18.15. Statistics
 - 18.16. Mass Comparison
 - 18.17. Differential weighing
 
 - 19. ACCESS TO BALANCE PARAMETERS
 - 20. TECHNICAL DATA
 - APPENDIX A – cooperation with EPSON printer
 - APPENDIX B – cooperation with KAFKA printer
 - APPENDIX C – cooperation with label printer CITIZEN
 - APPENDIX D – databases
 

- 63 - 
18.8.4. Parameters of parts counting mode – information 
Information items related to weighing are 
displayed in the grey coloured workspace area. 
The workspace enables containing up to 6 
pieces of information.  
If more than six parameters are enabled, then 
only first 6 are displayed. 
  Bargraph  YES   
A bargraph refers to operation of the filling and checkweighing modes in the 
parts counting mode.. The bargraph is a graphic interpretation of modes 
operation. The bargraph is visible only if the hot-key “Checkweighing” or 
“Filling” are enabled (set to YES) (set in menu for declaring presence of quick 
access keys) 
  Supplementary unit  YES   
This field contains measuring result expressed in an additionally determined 
measuring unit (supplementary unit) set in menu <Units>. 
  Tare  YES   
This field displays currently set tare value (displayed in currently set basic 
measuring unit). 
  Gross  YES   
This field displays current gross value (displayed in currently set basic 
measuring unit). 
  Counter  YES   
This field enables displaying data on number (sequence) of carried out 
measurements 
  Low threshold  NO   
The information indicating the value of low threshold that is declared for 
checkweighing mode, e.g. 20 pcs 
  High threshold  NO   
The information indicating the value of high threshold that is declared for 
checkweighing mode. 25 pcs 
  Target value  NO   
The information indicating number of parts to be added to counted parts to 
reach a target a in a filling process (in this case a filled quantity) 
  Statistics  NO   
  Displaying of the statistics component can be attributed as: Yes or No. 
  ID codes  NO   
“ID1“,“ID2“, “ID3“, “ID4“, “ID5“: Field enables determining identifiers set using 
a quick access key “ID“.  
Note: When selecting an ID value “ID1“, “ID2“, “ID3“ and “ID4“ the balance displays 
names (texts) corresponding to each determined ID. 