User manual

Table Of Contents
- 77 -
o low threshold
o high threshold
o target mass
o statistics /difference, amount, total sum, standard deviation, etc./
o ID codes
o non-standard data
o non-standard data project
Each information item features option YES or NO determining its displaying criterion.
CAUTION:
Defining non-standard information items is described in chapter 18.8.4.
Below description refers only to those buttons /pictograms/ that are not available in
working mode “Weighing”.
After enabling the working mode PERCENT
SETUP the software displays default set of
pictograms needed to immediate starting the
process:
set tare
reference mass
database of reference standards
set 100%
adjustment
Give reference mass /reference standard/
Pressing this hot-key opens a window with on-screen keyboard for entering
mass value, e.g. 350.0 g to be saved as a standard 100%. Based on this
standard, the balance carries out control process of the consecutive samples.
The mass value needs to have the
number of decimal places determined
PROCEDURE:
o Pres Quick access key
o The display shows a screen with
the reference mass
o Press field with the value of the
mass which opens a screen for
entering new mass value
o Accept by pressing OK button.
o Select number of decimal places.
o Pressing the button causes
automatic accepting the entered
value.
Checkweighing
Determine two checkweighing thresholds as values expressed in [%].