User manual
Table Of Contents
- 1. BASIC INFORMATION
 - 2. TRANSPORT AND STORAGE
 - 3. UNPACKING, INSTALLATION AND COMMISSIONING
 - 4. GETTING STARTED
 - 5. BALANCE LEVELING
 - 6. MENU STRUCTURE
 - 7. WEIGHING
 - 8. BASICS ON TERMINAL AND SOFTWARE OEPRATION
 - 9. BALANCE SOFTWARE
 - 10. BALANCE ADJUSTMENT
 - 11. DISPLAY CALIBRATION
 - 12. BALANCE USER SELECTION
 - 13. PERIPHERALS
 - 14. ABOUT
 - 15. COMMUNICATION
 - 16. UPDATE
 - 17. LEVELING
 - 18. BALANCE ADAPTATION – USER SETTINGS
- 18.1. Adding a new operator
 - 18.2. Deleting a operator
 - 18.3. Selecting a operator
 - 18.4. Operator settings
 - 18.5. Quick access to determining working modes parameters
 - 18.6. Parameter of working mode WEIGHING – quick access keys
 - 18.7. Parameter of working mode WEIGHING – information
 - 18.8. Parts Counting
 - 18.9. Checkweighing
 - 18.10. Dosing
 - 18.11. Percent setup
 - 18.12. Animal weighing
 - 18.13. Density of solids, liquid and air
 - 18.14. Formulations
 - 18.15. Statistics
 - 18.16. Mass Comparison
 - 18.17. Differential weighing
 
 - 19. ACCESS TO BALANCE PARAMETERS
 - 20. TECHNICAL DATA
 - APPENDIX A – cooperation with EPSON printer
 - APPENDIX B – cooperation with KAFKA printer
 - APPENDIX C – cooperation with label printer CITIZEN
 - APPENDIX D – databases
 

- 83 - 
5.  Mass of weighed load was increased to 
53,2g that is 106,4% of the reference 
mass value.  
The display status for this mass value 
is mass excess (bargraph colour is 
red). 
6.  – above the high threshold value. 
Caution; 
the tolerance of the high threshold is + 5% to 
the reference mass and the display indicates + 
6,6% thus the indication IS OUT OF the 
acceptance field. 
18.12. Animal weighing 
Animal weighing mode is intended to determining mass of an object 
that is moving on the weighing pan of a balance. Therefore, the 
structure of the working mode, i.e. its algorithm for determining stable 
mass indication. 
After enabling the animal weighing mode, it is immediately ready to be used. Follow 
the commands indicated in the COMMAND BAR of the display. After completing the 
weighing process, the measurement indication is locked on the display. Internal mode 
settings are editable using the QUICK access keys of the bottom display bar. 
The Quick access keys displayed in the bottom bar relating to the animal weighing 
mode. Set their attribute to Yes (present in the bottom bar) or No (absent).. 
Adjustment 
Clear counter 
Set tare 
ID definition 
Enable tare 
Animal weighing 
Print header 
Statistics 
Print footer  xxx  Non-standard buttons 
Principle of use and function of the quick access keys has been described in chapter 
18.6. of this user manual: Parameter of working mode WEIGHING – quick access 
keys. 
Other quick access keys function in accordance with the description provided in 
chapter 18.7. 