User manual
Table Of Contents
- 1. BASIC INFORMATION
 - 2. TRANSPORT AND STORAGE
 - 3. UNPACKING, INSTALLATION AND COMMISSIONING
 - 4. GETTING STARTED
 - 5. BALANCE LEVELING
 - 6. MENU STRUCTURE
 - 7. WEIGHING
 - 8. BASICS ON TERMINAL AND SOFTWARE OEPRATION
 - 9. BALANCE SOFTWARE
 - 10. BALANCE ADJUSTMENT
 - 11. DISPLAY CALIBRATION
 - 12. BALANCE USER SELECTION
 - 13. PERIPHERALS
 - 14. ABOUT
 - 15. COMMUNICATION
 - 16. UPDATE
 - 17. LEVELING
 - 18. BALANCE ADAPTATION – USER SETTINGS
- 18.1. Adding a new operator
 - 18.2. Deleting a operator
 - 18.3. Selecting a operator
 - 18.4. Operator settings
 - 18.5. Quick access to determining working modes parameters
 - 18.6. Parameter of working mode WEIGHING – quick access keys
 - 18.7. Parameter of working mode WEIGHING – information
 - 18.8. Parts Counting
 - 18.9. Checkweighing
 - 18.10. Dosing
 - 18.11. Percent setup
 - 18.12. Animal weighing
 - 18.13. Density of solids, liquid and air
 - 18.14. Formulations
 - 18.15. Statistics
 - 18.16. Mass Comparison
 - 18.17. Differential weighing
 
 - 19. ACCESS TO BALANCE PARAMETERS
 - 20. TECHNICAL DATA
 - APPENDIX A – cooperation with EPSON printer
 - APPENDIX B – cooperation with KAFKA printer
 - APPENDIX C – cooperation with label printer CITIZEN
 - APPENDIX D – databases
 

- 88 - 
o the selected density mode is 
automatically enabled on 
returning to weighing. 
Caution: after changing the working 
mode the default working mode is 
automatically changed  
/automatic change of the Density field 
value: e.g. liquids 
→
 solids/ 
18.13.4. Density of air 
Density of air is one of parameters required for calculating a correction value 
for a measuring result related to air BUOYANCY. Another parameter needed 
to calculate the correction value is density of tested sample. 
1.  The procedure requires a set of 
two standards: an aluminum one 
and a stainless steel one, both 
with determined mass. The set of 
weights is additional equipment 
of a balance 3Y series. 
PROCEDURE: 
o press Quick access key: AIR 
o to open a screen on density 
determining of air 
o press field with the value of the 
stainless steel standard and 
enter its value 
o press field with the value of the 
aluminum standard and enter its 
value 
o if the mass standards in use are made of other materials, then change 
the value of fields related to density of the material 
o press START field - the balance is ready to determine density of air 
o follow the commands displayed in the COMMAND BAR of the 
workspace area. 
− Place stainless steel standard on balance weighing pan 
− After stabilizing of the indication press OK button 
− Place aluminum standard on balance weighing pan 
− After stabilizing of the indication press OK button 
− Balance display shows calculated result of air density. 