User manual
38 
•  The program returns to weighing and the and the display shows  
the entered value with the „–” sign provided there was zero before  
on the display. 
13.5. Weighing for dual range scales 
Switching between the I range and the II range happens automatically 
(exceeding Max of the I range). 
Weighings in the second range is signalled by a pictogram in the top left 
corner of the display 
 . Then weighings is done with the accuracy  
of the II range to the moment of returning to zero (autozero range 
)  
where the scale switches back to the I range. 
Switching between the II range and I range is automatic both in the  
switching point the autozero zone. While in AUTOZERO
 – pictogram   
appears. Then pictogram 
 is off and a scale returns to weighing in the  
I range. 
13.6. Toggling between weighing units 
Operators can change the weight unit in two ways: 
•  Pressing the unit symbol on the screen, 
•  Pressing formerly defined button <
 Change unit>. 
Possible selection: 
•  gram [g] 
•  kilogram [kg] 
•  carat [ct] 
•  pound [lb] 
•  ounce [oz] * 
•  Newton [N] * 
*) – weighing unit inaccessible in a verified scale  