User Manual 4 load cell platform scales Manual number: ITKU-19-01-03-10-A Low profiled scales Mild steel scales Stainless steel scales Stainless steel scales (pit version) Ramp scales Pallet scales Beam scales MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS RADWAG 26 – 600 Radom Bracka 28 Street - POLAND Phone +48 48 38 48 800, phone/fax. +48 48 385 00 10 Selling department +48 48 366 80 06 www.radwag.
MARCH 2010 -2-
TABLE OF CONTENTS 1. INTENDED USE .......................................................................................................................................5 2. PRECAUTIONS ........................................................................................................................................6 2.1. Maintenance .....................................................................................................................................6 2.2. Accumulator / battery pack ....
15. WORK MODES ....................................................................................................................................40 15.1. Setting accessibility of operation modes.......................................................................................40 15.2. Selecting quantity of operation modes..........................................................................................41 15.3. Counting pieces of the same mass ...................................................
1. INTENDED USE Scales are designed for fast and precise measurements of weighed loads masses and direct commercial settlements. Tarring in full weighing range enables to determine net mass of weighed loads.
2. PRECAUTIONS 2.1. Maintenance A. Please, read carefully this user manual before and use the device according to its intended use. B. Devices that are to be withdrawn from usage should be sent back to the producer or in case of own utilization do it according to the law. 2.2. Accumulator / battery pack The device connected to mains inteligently monitors the battery state and charges it if possible.
Notice: Some symbols on accumulators identify harmful elements/compounds: Pb = lead, Cd = cadmium, Hg = mercury. 2.2.1. Power supply of weighing indicators in plastic casings Indicators in plastic casing are intended to be supplied from a power adapter or from NiMH rechargeable battery pack (standard equipment). New rechargeable batteries should be formatted according to the description in the chapter 14.4.4. of this manual.
• Remove discharged and then insert new batteries into the chamber, according to given polarity (+/-): • Close the lid of the chamber for batteries: In PUE C/31H and PUE C/31H/Z weighing indicators (stainless steel housing) the worn out accumulator can be exchanged to a new one by the authorized service of the manufacturer. 2.3. Operation in a strong electrostatic field If the device is about to operate in a strong electrostatic field (e.g. printing houses etc.) it should be connected to the earthing.
3. WARRANTY CONDITIONS A. RADWAG is obliged to repair or change those elements that appears to be faulty because of production and construction reason, B. Defining defects of unclear origin and outlining methods of elimination can be settled only in participation of a user and the manufacturer representatives, C. RADWAG does not take any responsibility connected with destructions or losses derives from non-authorized or inappropriate (not adequate to manuals) production or service procedures, D.
4. MAIN DIMENSIONS 4.1.
4.2. Ramp scales 4.3.
4.4. Stainless steel scales (pit version) 4.5.
4.6. Pallet scales 4.7.
5. ASSEMBLY AND LEVELLING 5.1. Assembly of scales 5.1.1. 4 load cell platform scales Before installing remove the transport protection. Screw in levelling feet to mandrels placed in load cells 5.1.2. Ramp scales Before setting the scale mount the steel camping ring (2) to the basis of feet (1) using expanding ring (3) and then screw in the basis of the foot (1) in the mandrel (4). Put the scale on the flat ground. Put ramps on steel clampings.
5 - nut 6 - washer 7 – centring washer 8 - anchor 5.2. Levelling the scale Level scale by means of turning levelling feet and external level indicator. Put the pads under the regulation legs and observe the level condition indicator. The instrument is levelled correctly if the air bubble do not move more than 2mm from the central position. Each leg can be screwed in and out in order to level the scale. Correct levelling can be done by putting steel pads under the legs. 6.
7. KEYPAD 8. KEYS’ FUNCTIONS Switching on/off Function key (operation mode selection) Sending a weighing result to RS232 Zeroing Tarring Notice: + keys’ functions changes. The way of operation After pressing in this mode is described in details further in this manual.
9. INSCRIPTIONS ON THE DISPLAY No Text string Description 1. FIL Filter level 2. bAud Transmission baud rate 3. PCS Piece counting 4. HiLo +/- control according to a standard mass 5. rEPL Automatic printout 6. StAb The condition of printing data 7. Auto Autozero correction 8. t1 Power save – time to switch off while no operation 9. toP Latch of the max measurement 10. Add Totalizing 11. AnLS Weighing animals 12. tArE Memory of 9 tare values 13.
10. USER MENU 10.1. Submenus User’s menu is divided into 6 basic submenus. Each group has its own characteristic name preceded by the letter P and a number. P1 rEAd P 1.1 Fil P 1.2 Auto P 1.3 tArA P 1.4 Fnnd P2 Prnt P2.1 Pr_n P2.2 S_Lo P2.3 bAud P2.4 S_rS P3 Unit P3.1 StUn P4 Func P4.1 FFun P4.2 Funi P4.3 PcS P4.4 HiLo P4.5 PrcA P4.6 Prcb P4.7 AtAr P4.8 toP P4.9 Add P4.A AnLS P4.b tArE P5 othr P5.1 bL P5.2 bLbt P5.3 bEEP P5.4 t1 P5.5 CHr6 P6 CAL P6.1 St_u P6.
10.2. Browsing user menu Use scale’s keys to move inside the menu. 10.2.1. Keypad Entering main menu + + Inscribing tare value Increasing a digit value by „1” moving down in the menu Battery / accumulator state monitoring + Toggling between gross / net values + Selecting the parameter or changing the value of a selected parameter Entering the selected submenu or activating a parameter for changes Confirmation (enter) Leaving without changes or reaching a higher level of the menu 10.2.2.
11. WEIGHING Put a load you want to weigh on the weighing pan. When the pictogram appears it means that the result is stable and ready to read.
• For four load cell platform scales (pit version) – loads carried and weighed together with trucks: − Total weight of the truck and its load cannot exceed the maximum value of the scale range, − Trucks are allowed to drive through without braking and accelerating with the maximum speed 2km/h, − Platforms should be driven over only perpendicular to the edge of the platform, in order to uniformly load the scale, − Loads should be stand steadily and slowly, − It is not permitted to push or pull loads on or fr
Notice: Tarring cannot be performer when a negative or zero value is being displayed. In such case appears on the display and short audible signal will be emitted. 11.2. Inscribing tare value You can also inscribe a tare value. While in weighings mode press: • Press simultaneously • You will see : and , • Using • • , Press Program returns to weighings mode. The inscribed tare value can be seen on the display with „–” sign, Tare can be inscribed anytime in weighings mode.
11.3. Zeroing To ZERO the scale press: . The scale will display zero and following pictograms: and . Zeroing is only possible within the scope of ±2% of full scale. While zeroing outside the scope of ±2% you will see . Zeroing is possible only in stable state. Notice: Zeroing is possible only within the ±2% interval of the maximal range. If zeroing is performed beyond this range the message and short audible signal will be emitted. 11.4.
• press , until the expected unit appears on the display: Options: A. When the basic unit is [kg], users can toggle between: [kg, lb, N], for verified scales [lb] is not accessible, B. If the basic unit is [g], users can toggle between: [g, ct, lb], for verified scales [lb] is not accessible, • After you select the unit press , the scale returns to: • Return to weighing according to chapter - 10.2.2. Notice: After turning on the scale always sets the basic unit.
11.6. Temporarily selected unit This function is used to set weight unit the scale will use temporarily until the next power off or next selection. Procedure: • Press and then: • After you select the unit you want come back to weighing procedure. Options: A. When [kg] is a basic unit, users can select following units: [kg, lb, N], [lb] is not accessible for verified scales. B. When [g] is a basic unit, users can select following units: [g, ct, lb], [lb] is not accessible for verified scales. 12.
12.1. Setting a filtering level Procedure: • Enter the submenu and then: 1 - 4 - level of filtering • By pressing select the filtering level you need Notice: Filtering level influences the time of stabilization. The higher the filtering level is the longer stabilization time is needed. Return to weighing: See - 10.2.2. 12.2. Median filter This filter eliminates short changes (impulses) of measure signal (e.g. shocks). Procedure: • Enter the submenu
Fnnd Fnnd no - filter disabled YES - filter enabled Return to weighing: See - 10.2.2. 12.3. Autozero function The autozero function has been implemented in order to assure precise indications. This function controls and corrects „0” indication. While the function is active it compares the results continuously with constant frequency. If two sequentional results differ less than the declared value of autozero range, so the scale will be automatically zeroed and will be displayed.
Procedure: • Enter the submenu and then: Fnnd Fnnd no - filter disabled YES - filter enabled Return to weighing: See - 10.2.2. 12.4. Tare function This parameters enables users to configure a tare function. Procedure: • Enter the submenu
tArA AtAr - automatic tare function on and is stored in balance memory after unplugging it from mains (Description of function operating point 15.6 automatic tare) tArA no - automatic tare function off (user can turn on operating of automatic tare F6 AtAr – till unplugging the balance from mains) tArA tArF - tare memory function – stores last value of tare in balance memory. It is automatically displayed after starting the balance.
13. RS 232 PARAMETERS External devices connected to RS 232C have to be supplied from the same mains and common electric shock protection. It prevents from appearing a potential difference between zero leads of the two devices. This notice does not apply to the devices that do not use zero leads. Transmission parameters: • • • • Baud rate - 2400 – 38400 bit / s Data bits - 7,8 Stop bits - 1,2 Parity control - no, even, odd.
Pr_n noStAb - Pr_n Pr_n Pr_n Pr_n StAb rEPL CntA Cntb - immediate printout (not accessible in verified scales) sending stable results automatic operation continuous transmission in basic unit continuous transmission in present unit Return to weighing: see 10.2.2. 13.2. Minimal mass threshold This function is necessary while working with automatic tare or automatic operation or weighing animals.
Return to weighing: see 10.2.2. 13.3. Baud rate Procedure: • Enter the submenu
Return to weighing: see 10.2.2. 13.4. Serial transmission parameters Procedure: • Enter the submenu
14. OTHER PARAMETERS The user can set parameters which influence the scale operation. They are gathered in the submenu e.g. backlight and beep signal. Enter this submenu according to chapter 10.2. 14.1. Backlight function Program recognises the way the scale is supplied (mains, battery) and automatically selects the way of operating on the backlight: bl blbt – for mains, – for batteries or rechargeable battery pack. 14.1.1.
Return to weighing: See 10.2.2. Notice: When bl=Auto, and the indication has not changed for 10s, the backlight is automatically switched off. The backlight is switched on again automatically after the result changes. 14.1.2. Backlight for supplying from batteries The user can change the intensity of backlight from 0% to 100%. The lower the intensity is the longer the scale operates without recharging or exchanging batteries. When the intensity is set this function works as AUTO (described above).
14.2. “Beep” signal – after pressing a key Procedure: • Enter the submenu and then: bEEP bEEP no - switched off YES - switched on Return to weighing: See 10.2.2. 14.3. Automatic switch-off This function is essential to save the battery power. The scale is switched off automatically when (function t1 = YES) no weighing appears in 5 minutes. (no changes on the display). In case when this function disrupts the operation (e.g.
* automatic enabling/disabling according to the source of power. Procedure: • Enter the submenu and then: Returnto weighing: See 10.2.2. 14.4. Battery voltage level check While supplying from batteries too low level of voltage is measured by software the pictogram is displayed. It means that charging or exchanging batteries is required. 14.4.1. Checking the batteries This function is to check the level of battery supply.
Procedure: After displaying the level of batteries (in per cents) the program returns to weighing. 14.4.2. Battery discharge pictogram The symbol (bat low) switches on when the voltage level drops to 18% of the accepted level of voltage. It means that charging or exchanging batteries is required.
• • Message appears on the display (for indicators in plastic casings) or (for indicators in metal housings). In case of damaging accumulators or lack of it the pictogram blinks quickly (period about 0.5 sec). Notice: Indicators in plastic casings are equipped with the set of rechargeable batteries NiMH R6 (AA) and power adapter. Procedure: • Enter the submenu and then: CHr6 YES CHr6 no - enabled - disabled Return to weighing: See 10.2.2. 14.4.4.
Procedure: 1. 2. 3. 4. 5. Supply the indicator from mains. Charge batteries for 12 hours (time of charging 2200mAh batteries). After 12 hours unplug from mains. Use the device up to the moment of self powering down. Repeat the process of charging starting from point 1. Notice: They reach their optima capacity after three cycles of full charging and discharging. 15. WORK MODES 15.1.
no YES – mode is disabled – mode is enabled Return to weighing: See 10.2.2. 15.2. Selecting quantity of operation modes This function enables user to set if ,after pressing key, all operating modes will be accessible (ALL) or only one from the list chosen and used by operator. Procedure: • Enter the submenu and then: After choosing setting press key. The program will return to displaying name of submenu . Return to weighing: See 10.2.2.
15.3. Counting pieces of the same mass Standard solution is equipped with option of counting small pieces of the same mass. It is possible to execute a tare function in this operating mode in order to tare a container value. Notice: 1. Counting pieces does not work together with other scale functions, 2. The counting pieces function is not saved as a default start function so it is not remembered after restarting. Procedure: • Enter to function: • You will see a blinking value of sample quantity.
• If option is choosen in the scale program displays estimated unit mass of the last piece (about 3 sekonds) and then goes to Counting pieces automatically setting the previously displayed value as valid for the procedure.
Notice: key when load is not present on the pan, 1. If a user presses the the message -Lo- will be indicated for a few seconds and the scale will automatically return to weighing. 2. In order to comply with the rules of appropriate counting pieces put as many pieces as possible during unit mass adjustment. Single piece mass should not be less than 5 divisions. 3. If a single piece mass is lower than a reading interval d the display will show the message (see ch. 20.
• The inscribed value confirm by pressing , the program will automatically go to the higher threshold of weighing (Max): • The inscribed value confirm by pressing automatically go to the main window. , the program will • During setting threshold values following cases take place: Notice: If a user erroneously enters a value of the lower threshold higher than the upper one, the scale will indicate an error message and will return to weighing.
Return to weighing: • Press the key twice. 15.5. Control of % deviation referring to the inscribed standard mass Scale software enables control of deviation (in %) of weighed loads mass referring to the inscribed standard mass. Mass of standard can be determined by its weighing (PrcA function) or entered to the scale memory by an user (PrcB function). 15.5.1.
Return to weighing: • Press the key twice. 15.5.2. Mass of standard inscribed to scale memory Procedure: • Enter to function: • The program goes to the weight display window: • Using where: and set standard mass, - digit selection, - digit setting. , • Confirm the entered value by pressing • You will see the indication equal to 0,000%, • From this moment display will not indicate the mass of weighed load but deviation of the load mass placed on the pan referring mass of standard (in %).
Return to weighing: • Press the key twice. 15.6. Automatic tare This function is useful for fast net mass determination of weighed load in case when tare value of is different for each load.
15.7. Measurement max force on the pan – latch Procedure: • Enter to function: • Confirmation of choice of function is indication of the Max pictogram: • Apply a force to the weighing pan, • The display of scale will latch the maximum value of the force, • Remove loads from the pan • Before the next measurement press the key. Return to weighing: • Press the key twice. 15.8. Totalizing Scale software is equipped in a totalizing function of single weighings.
15.8.1. Enabling the work mode Procedure: • Enter to function: • A letter „P” in the left side of the display is a confirmation that function have been selected: 15.8.2. Totalizing procedure • Enter function according to ch. 15.8.1, • Put the first load on the pan. If the weighing procedure is performed in a container put the container on the pan first and tare it.
• Take off the load from the pan, indication returns to ZERO and the letter „P” in the left part of the display appears, • Put the next load on the pan, , the sum of first and second weighing • After stabilizing press will appear on the display, the „▲” pictogram in the upper right corner will be displayed and the second weighing result will be printed on the printer connected to the indicator: to complete the procedure (with the loaded or unloaded • Press pan), a sum of all weighings will be printed: (1)
- In order to continue the procedure press , the indication returns to ZERO and the letter „P” appears in the left part of the display. The scale is ready for weighing. - In order to terminate the previous totalizing procedure press , or . You will see the letter „P” in the left part key, of the display. The scale is ready for weighing. 15.8.4.
15.9.
• Load an animal to the platform, • After exceeding the -LO- value (see 13.2), program starts the weighings process.
• The program goes to displaying the first value from the selection of tare values (press to chose different values): • After selecting the right position press field: and you will see an editing • Enter the selected tare value to the scale memory - 55 - ,
• The program returns to the following window: Return to weighing: • Press . 15.10.2. Selecting a tare value from the memory • Enter function according to the ch. 15.10.
16. USER CALIBRATION Only for non-verified scales Confirmation of high accuracy of weighing requires periodical correcting of calibration factors in the scale memory – this is adjustment of the scale. Calibration should be performed when we start weighing or dynamic change of temperature occurs. Before starting calibration remove loads from the pan. 16.1. Calibration Procedure: • Enter the submenu
• Put a weight of the displayed mass value on the pan and press . The calibration process will start which is signalled by the message: • After completion of the process of calibration the following screen will appear • Take off the weight , then the following sequence of screens will appear • Calibration process can be terminated anytime by pressing which is signalled by the following message on the display: • Return to weighing with saving changes that have been made.
16.2. Start mass adjustment If the scale does not require the full calibration process sit is possible to adjust only a new start mass. Procedure: • Enter the submenu
• Return to weighing performing the procedure of saving parameters. Caution: If the start mass adjustment lasts longer than 15 the message will be displayed and short audible signal will be emitted. Press to perform calibration again with more stable ambient conditions! 17. COOPERATION WITH PRINTER key is pressed a current mass value together with Each time the mass units is sent to RS 232 interface. Depending on setting of STAB parameter it can be printed out with temporary or stable value.
Cable diagrams: Scale – Kafka printer cable diagram for plastic casing Scale – Kafka printer cable diagram for steel housing 18. COOPERATION WITH COMPUTER Sending weighing results to the computer can be done: - manually - in continuous way - - automatically - on the request from the computer - key, after pressing after function activating or sending an appropriate command, After stabilizing the indication - After sending a control command These scales can cooperate with „EDYTOR WAG” program.
Cable diagrams: Scale – computer cable diagram for plastic casing Scale – computer cable diagram for metal housing 19. COMMUNICATION PROTOCOL 19.1. General information A. A character protocol scale-terminal has been designed for communication between RADWAG scales and external devices via RS-232 interface. B. It consists of commands sent from an external device to the scale and a responses from a scale. C. Responses are sent every time after receiving a command (reaction for any command). D.
19.2.
XX _ - command name - substitutes spaces 19.4. Command’s description 19.4.1. Zeroing Syntax Z CR LF Possible answers: Z_A CR LF Z_D CR LF - command accepted and in progress - command completed Z_A CR LF Z_^ CR LF - command accepted and in progress - command comprehended but zero range overflow appeared Z_A CR LF Z_E CR LF - command accepted and in progress - time limit for stable result exceeded Z_I CR LF - command comprehended but cannot be executed 19.4.2.
TO_TARA CR LF - command executed Frame format: 1 2 3 4 5-6 7-15 16 T O space stability space tare space Tare Unit 17 18 19 unit 20 21 CR LF - 9 characters with decimal point justified to the right - 3 characters justified to the left 19.4.4.
SI_I CR LF - command comprehended but cannot be executed at the moment MASS FRAME - mass value in basic unit is returned Frame format: 1 2 3 4 5 6 7-15 16 S I space stability space sign mass space 17 18 19 unit 20 21 CR LF Example: S I CR LF – computer command S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF - command done, mass value in basic unit is returned immediately. 19.4.6.
19.4.7. Send the result immediately in current unit Syntax: SUI CR LF Possible answers: SUI_I CR LF - command comprehended but cannot be executed MASS FRAME - mass value in current unit is returned immediately Frame format: 1 2 3 4 5 6 7-15 16 S U I stability space sign mass space 17 18 19 unit 20 21 CR LF Example: S U I CR LF – computer command S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF - command executed and mass returned 19.4.8.
Possible answers: C0_I CR LF - command comprehended but cannot be executed C0_A CR LF - command comprehended and executed 19.4.10. Switch on continuous transmission in current unit Syntax: CU1 CR LF Possible answers: CU1_I CR LF - command comprehended but cannot be executed CU1_A CR LF MASS FRAME - command comprehended and in progress - mass value in current unit is returned Frame format: 1 2 3 4 5 6 7-15 16 S U I stability space sign mass space 17 18 19 unit 20 21 CR LF 19.
19.5. Manual printouts / automatic printouts Users can general manual or automatic printouts from the scale. • Manual printouts can be performed after loading the pan . and stabilizing indication by pressing • Automatic printouts can be performed only after loading the pan and stabilizing indication. Notice: If a scale is verified printouts of immediate values are blocked.
19.6. Continuous transmission The indicator can work in a continuous transmission mode. It can be switched on or off in parameters or using RS232 commands. The frame format sent by the indicator in case of setting
20.
21. TECHNICAL PARAMETERS 21.1.
Scale type: WTC /4 3000 C10 WTC /4 3000 C11 WTC /4 6000 C11 WTC/4 12000 C14 Max capacity 3000kg 6000kg 12000kg Min capacity 20kg 40kg 100kg Readability 1kg 2kg 5kg Verification division 1kg 2kg 5kg -6000kg -12000kg Range of tare -3000kg Operation temperature -10°C to +40°C Ingress protection rating IP 65 steel, IP 43 plastic Power supply Supplied from batteries 230V AC, 50Hz / 11V AC, and 6×AA (NiMH) Average operation when supplied from batteries 35h Display LCD with backligh
Scale type: WPT/4 1500 H8 Max capacity WPT/4 1500 H8/9 WPT/4 1500 H9 WPT/4 3000 H8 1500kg 3000kg Min capacity 10kg 20kg Readability 0,5kg 1kg Verification division 0,5kg 1kg -1500kg -3000kg Range of tare Operation temperature -10°C to +40°C Ingress protection rating IP 68 platform, IP 66/67 indicator 230V AC, 50Hz / 11V AC and internal gell cell SLA 6V accumulator Power supply Supplied from batteries Average operation when supplied from batteries 45h Display LCD with backlight Bri
21.3.
Scale type: WPT/4 3000 H8/9/Z WPT/4 3000 H9/Z WPT/4 3000 H10/Z WPT/4 6000 H10/Z Max capacity 3000kg 6000kg Min capacity 20kg 40kg Readability 1kg 2kg Verification division 1kg 2kg -3000kg -6000kg Range of tare Operation temperature -10°C to +40°C Ingress protection rating IP 68 platform, IP 66/67 indicator 230V AC, 50Hz / 11V AC and internal gell cell SLA 6V accumulator Power supply Supplied from batteries Average operation when supplied from batteries 45h Display LCD with backligh
WPT/4N 1500H2 Scale type: WPT/4N 1500H3 Max capacity WPT/4N 1500H4 1500kg Min capacity 10kg Readability 500g Verification division 500g Range of tare -1500kg Operation temperature -10°C to +40°C Ingress protection rating IP 68 platform, IP 66/67 indicator 230V AC, 50Hz / 11V AC and internal gell cell SLA 6V accumulator Power supply Supplied from batteries Average operation when supplied from batteries 45h Display LCD with backlight Platform size 1,1 x 1,2m 1,2 x 1,5m 190kg 230kg 2
Scale type: Max capacity Min capacity Readability Verification division Range of tare Operation temperature Ingress protection rating Power supply Supplied from batteries Display Platform height * Bridge dimension [m] Inroad dimension [mm] Weight of scale with indicator Inroad weight WPT/4F 150 C7 WPT/4F 300 C6 WPT/4F 300 C7 WPT/4F 300 C8 150kg 1kg 50g 50g -150kg 300kg 2kg 100g 100g -300kg -10°C to +40°C IP 65 steel, IP 43 plastic 230V AC, 50Hz / 11V AC, and 6×AA (NiMH) Average operation when supplied
21.6.
21.8. Beam scales Scale type: WPT/4P2 600 C WPT/4P2 1000 C WPT/4P2 2000 C WPT/4P2 3000 C Max capacity 600kg 1000kg 2000kg 3000kg 4kg 10kg 20kg 20kg 0,2kg 0,5kg 1kg 1kg Min capacity Readability Verification interval Tare range 0,2kg 0,5kg 1kg 1kg -600kg -1000kg -2000kg -3000kg Beam length 1,2m Max.
Scale type: WPT/4P2 2000 C2 WPT/4P2 4000 C2 WPT/4P2 6000 C2 Max capacity 2000kg 4000kg 1000kg Min capacity 20kg 40kg 40kg Readability 1kg 2kg 2kg Verification interval Tare range 1kg 2kg 2kg -2000kg -4000kg -6000kg Beam length 2,5m Max.
Scale type: WPT/4P2 2000 H1 WPT/4P2 4000 H1 WPT/4P2 6000 H1 Max capacity 2000kg 4000kg 1000kg Min capacity 20kg 40kg 40kg Readability 1kg 2kg 2kg Verification interval Tare range 1kg 2kg 2kg -2000kg -4000kg -6000kg Beam length 2m Max.
22. TROUBLE SHOOTING Problem Turning on does not work The scale turns off automatically Cause Solution Discharged batteries. Connect to mains or change batteries No batteries (not installed or improperly installed) Check the correctness of installation (polarization) „t1” set to „YES” (Power save) In „othr” submenu change „5.4 t1” to „no” After turning on „LH” Loaded weight pan during message on the display powering up Unload the pan. Then the scale will indicator zero. 23.
• • • • Handle for measuring indicator in plastic version, A rack for PUE C/31, PUE C/31H or PUE C/31H/Z indicator, Anti-dust case for Epson printer, Inroads for low profile scales. Computer programs: • “EDYTOR WAG” computer program, • "RAD-KEY" computer program, • "PW-WIN" computer program.
MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS RADWAG WAGI ELEKTRONICZNE 26 – 600 Radom, Bracka 28 Street POLAND Tel. +48 48 38 48 800, tel./fax. + 48 48 385 00 10 Selling departament + 48 48 366 80 06 www.radwag.