Technical data
40 
11.4.2.  External calibration 
The external calibration should be performed with external mass class F
2
•  Start external calibration process 
balances with external calibration 
•  Order to load off the pan appears on the display (no load on the pan). After 
yesing load off the pan press the ENTER key. 
•  The balance determines mass of empty pan 
•  Put load and press the ENTER 
•  After the calibration the balance returns to submenu P1.2 uCAL 
•  Return to weighing – as in the point 5.1.2. 
If balance is verified user does not have the possibility to perform external 
calibration process. 
11.5. Calibration report printout 
After cali
bration user can receive the calibration report. The report can be 
printed on connected printer and sent to computer or recorded in file. 
P1.6 CALr:  no  – report is not printed 
P1.6 CALr:  YES  – report is printed 
If the parameter has the value YES, the report is generated and sent 
automatically. 