Technical data
68 
15.2. Printing data with date and time 
Every printout of weighing can be printed with date and time of measurement.  
It’s possible if balance is connected to printer Kafka 1/Z or Kafka SQ S. 
After connecting balance and printer set in balance menu parameters of P2 
GLP group:  
PdAt on value YES 
Ptin on value YES 
15.3. Cooperation with statistic printout Kafka SQS 
After conn
ecting balance to printer KAFKA SQ S statistic of measurement is 
possible to do. Example of printout with statistics from series of measurement: 
1 9:02:15  + 7.0016  g 
2 9:02:39  + 5.0152  g 
3 9:02:58  + 
12.0171 g 
4 9:03:15  + 9.9937  g 
5 9:03:34  + 12.0169  g 
6 9:03:48  + 22.0111  g 
Data 13.09.2001 Godz. 9:04 
n     6      batch quantity 
sum x    68.0556  g  summary of mass of samples 
x     11.34260  g  average value 
s    5.92328 g  standard deviation 
srel   52.22    %  factor of variancy 
min    5.0152    g  min value 
max    22.0111  g  max value 
R    16.9959  g  difference max – min 
15.4. Format of sending data 
Result of 
weighing can be sent from balance to external device after pressing 
the PRINT button on the balance or after sending order from the computer. 