Technical data
69 
15.4.1.  Format of sent data after pressing the PRINT button 
Depending on setting of P4.5 PStb parameter only stable measurement or 
mass of instantaneous mass will be sent. 
For verified balance printout of temporary measurements will be blocked 
for instantaneous measurements. 
Format of printout 
1  2  3  4 - 12 13 14 - 16 17 18
stability 
marker 
space 
mark 
mass 
space 
unit 
CR 
LF 
stability marker – [space] if stable  
[?] if unstable 
[^] if there is an error of exceeding the range on + 
[v] if there is an error of exceeding the range on - 
Mark  – [space] for positive values or [-] for negative values 
mass  – 9 marks alignment to the right 
units  – 3 marks alignment to the left 
15.4.2.  Format of sent data for orders generated from the computer 
Bala
nce after receiving the command answers first: 
XX_A  CR LF  – commend understood, realization started  
XX_I CR LF  – commend understood, but unavailable at the moment 
XX _ ^ CR LF  – commend understood, but max range is exceeded 
XX _ v CR LF  – commend understood, but min range is exceeded 
XX _ E CR LF   – error occurred while command realization – exceeded time limit 
while waiting for stable result (time limit is characteristic 
parameter of balance) 
XX    – name of command 
And afterwards 