7.  START-UP 
When plugged to mains, the balance displays name and number of software, and 
afterwards proceeds to weighing mode.  
7.1 Balance temperature stabilization period 
Before start of measuring processes, it is necessary to wait until the balance is 
thermally stabilized. It is a period of so called balance self-heating. In case of 
balances PS series self-heating period takes approximately 30 minutes. The 
specified time interval refers to balances that have been stored in room temperature 
before plugging to mains.
For balances that were stored in much lower temperatures before plugging to mains 
(e.g. during winter period) thermal stabilization should last approximately 4 hours. 
Correct operation of a balance is possible within temperature range specified in the 
technical parameters, see point 1. If the balance is placed in area which ambient 
temperature is out of temperature range specified in the technical data, then 
balance’s display may signal such situation by lighting an appropriate marker (a 
black triangle). 
In case the balance displays the marker, increase or decrease ambient 
temperature, depending on present temperature value. It is recommended that 
ambient temperature changes at balance’s place of use are very small (slow). 