9.  WEIGHING 
Before start of weighing process or in case of essential change of ambient 
conditions at a workstation (e.g. ambient temperature change at a workstation more 
than 3
o
C) the balance requires adjusting. the procedure of balance adjustment is 
described further in this user manual.  
•  Before start of weighing procedure, it is recommended to load the balance’s 
weighing pan a few times with mass close to balance max capacity, 
•  Check if unloaded balance indicates “precise zero” - 
 and whether 
measurement is stable - 
, if not press ESC/TARE key; 
•  Press Units key to set a measuring unit:  
[g], [mg], [ct], and also if enabled in factory menu [kg], [lb], [oz], [ozt], [dwt], 
[t], [mom], [gn], 
•  place weighed object on balance’s weighing pan and read result only on 
stabilization of measurement result, 
•  mass indication of a load placed on balance’s weighing pan can be tarred for 
multiple times by pressing ESC/TARE  key ((pay attention not to exceed 
maximal capacity of a balance by applying multiple tare function). 
During times between carrying out the following measurement series do not unplug 
the balance from mains. It is recommended to switch off  balance’s display by 
pressing  ON/OFF  key. On repeated pressing of the ON/OFF  key the balance is 
ready for operation and carrying out the following measurements. 