Technical data
42 
adjustment is automatically initiated if temperature changes more than 3
o
C 
from last saved temperature.  
-  adjustment triggered by elapsing time. It is possible to declare time 
intervals which are a criteria for balance adjustment. Accessible settings 
are: adjustment after 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12 hours since last 
carried out adjustment procedure. 
CAUTION 
Declaring time interval for balance adjustment is available only in 
non-verified balances. In case of verified balances the time interval 
between following adjustments is set for 3 hours 
12.2.  Menu adjustments settings 
P1 CAL 
P1.1  iCAL    |   
P1.2  uCAL    |   
P1.3  tCAL    |   
P1.4  ACAL    |  both 
P1.5  CALt    |  1 h 
P1.6  CALr    |  YES 