Technical data
43 
•  P1.1 iCAL – Internal adjustment 
Start of internal adjustment process. Adjustment is conducted automatically, and 
no operator assistance is required. If weighing pan of the balance is loaded, 
balance will display a command to unload the weighing pan. If weighing pan is 
loaded with relativaly small mass (up to 4 % of balance max capacity) than 
adjustment process will be carried out automatically, but measuring results may 
differ by the amount that was present on balance weighing pan. 
 or CAL function key 
P1.2 ECAL – External adjustment (with an external weight) 
adjustment with external weight, which value is saved in factory menu of 
balance, function unavailable in verified balances. 
balances with internal adjustment  balances with external adjustment 
P1.3 tCAL – Adjustment test  
mass comparison of internal adjustment weight with its value saved in balance 
memory. 
balances with internal adjustment   balances with external adjustment 