Technical data
81 
16.4. Format of data sending 
Weighing result can be sent to a peripheral device by pressing PRINT key on a 
balance keyboard or by setting a control command from computer level.  
16.5.  Format of data sent on pressing of PRINT key 
Depending on setting of P4.5 PStb parameter only stable measurement or 
instantaneous mass indication will be sent.  
CAUTION: 
For verified balances printout of instantaneous measurement data is 
blocked.  
Printout format  
1 
2 
3 
4 - 12 
13 
14 - 16 
17 
18 
stability 
marker 
space 
sign 
mass 
space 
unit 
CR 
LF 
stability marker – [space] if stable  
-  [?] if unstable 
-  [^] if there is an error of range exceeding on + 
-  [v] if there is an error of range exceeding on - 
sign   – [space] for positive values or [-] for negative values 
mass  – 9 marks alignment to the right  
unit   – 3 marks alignment to the left 