Technical data

82
16.5.1. Format of data sent as response for commands generated
from a computer
On receipt of a command, balance responses with:
XX_A CR LF command understood, in progress
XX_I CR LF command understood, but currently not available
XX _ ^ CR LF command understood, but max range is exceeded
XX _ v CR LF command understood, but min range is exceeded
XX _ E CR LF error occurred while carrying out the command time limit
exceeded while waiting for stable measurement result (time limit
is characteristic parameter of balance)
XX name of command
And followed by:
1 - 3
4
5
6
7
8 - 16
17
18 - 20
21
22
Command
space
stability
marker
space
sign
mass
space
unit
CR
LF
Command 1 ÷ 3 characters
stability marker [space] if stable
[?] if unstable
[^] if there is an error of range exceeding on +
[v] if there is an error of range exceeding on -
sign [space] for positive values or [-] for negative values
mass 9 marks alignment to the right
unit 3 marks alignment to the left