User manual
- 123 -
34.2.1. Technical specification 
Parameters of outputs 
Quantity of outputs  8 
Type of outputs  Reed operation contacts 
Wire diameter  0,14 - 0,5mm
2
Maximal load-current contact capacity  0,2A DC 
Maximal forward voltage  50V DC 
Parameters of inputs 
Quantity of inputs  8 
Input type  Optoinsulated 
Wire diameter  0,14 – 0,5mm
2
Control voltage range  5 -24V DC 
34.2.2. Installing method in PUE 5 terminal 
Module is designated for assembly inside terminal PUE5. Module 
is assembled to main board of terminal to 10-pin J22 connector. For 
module WE8 gland is installed on casing lid, 3m cable ended with 
insulated conductors is led through gland. 
Installing procedure: 
1.  Unplug the terminal (remove the plug from socket 230V); 
2.  Unscrew the casing lid (back part of terminal casing); 
3.  Install the module in connector J22 on main board; 
4.  While installing module pay attention to plastic posts fastening 
to main board. They should be placed in assembly holes in main 
board and in assembly holes in module WE 8, 
5.  In one of free glands remove the stopper and led through it cable 
IN/OUT (if necessary use bigger gland depending on cable diameter); 
6.  Connect the cable IN/OUT to joint J1, J2 for outputs and J3, J4 for 
inputs on module 8IN/8OUT according to description given in table; 
7.  Connect cable IN/OUT to group of conductors (undo band clips 
fastening group of conductors, lay the cable and clamp band clips – 
multiple use band clips); 
8.  Assembly cover of indicator casing. 