37.  APPENDIX A – SETTING A BARCODE SCANNER 
1.  RADWAG scales use RS232 interface with transmission in one 
direction to communicate with a barcode scanner. It requires only two 
wire connection. So barcode scanners needs to be equipped with 
RS232 with disabled hardware and software flow control. 
2.  Both the terminal and a scanner can have transmission parameters 
set. Both devices should have set the same parameters: baud rate, 
data bits, parity control, number of stop bits. e.g. 9600,8,N,1 – baud 
rate 9600 bit/s , 8 - data bits, no parity control, 1 stop bit. By default in 
PUE5 barcode scanner needs to be connected to COM3 – with baud 
rate 9600 bit/s. 
3.  Barcode scanners can send additional information apart from a 
barcode e.g. barcode symbology. As RADWAG devices do not use 
this information it is advisable to disable it. 
4.  Some RADWAG systems can omitted some inessential information 
in the code by setting parameters outlining the beginning and length 
of the analyzed code. 
5.  In order to have a barcode read by PUE5 it is required to program an 
appropriate prefix and suffix. The prefix is (1 byte) 01 hexadecimally 
and the suffix is (1 byte) 0D hexadecimally. 
6.  Different barcode scanners allow to disable/enable reading different 
barcode symbologies. 
7.  Barcode scanners can be programmed by reading different 
programming codes present in their programming manuals. 
8.  Barcode scanners bought together with RADWAG systems are 
properly configured and tested. 
Barcode coded hexadecimally with prefix 
and suffix 
Barcode in ASCII code 
(omitted control 
characters) 
Symbology 
01 30 30 32 31 30 31 32 36 0D  00210126  EAN-8 
01 30 31 32 33 34 35 36 37 38 39 0D  0123456789   2 of 5 
01 43 4F 44 45 20 33 39 20 54 45 53 54 0D  CODE 39 TEST  CODE 39 
01 31 31 30 31 32 33 34 35 36 37 38 39 31 0D  1101234567891  EAN-13 
01 43 6F 64 65 20 31 32 38 20 54 65 73 74 0D  CODE 128 Test  CODE 128 