Manual

In the overlap <Transmission settings>:
Set the communication mode as „RS232 transmission”,
Select an appropriate com number,
Configure the selected port (baud rate, parity, data bits, stop bits),
Confirm the changes by clicking ,
Rerun the program,
Press the „Connect” button,
The appropriate communication with scales is signalled by pictogram
in the left part of the display,
Transmission problems are signalled by the message window:
32.2.3. Setting Ethernet parameters
On order to establish the connection with EDYTOR WPW through
Ethernet interface follow the remarks below:
Power up the scale and run the EDYTOR WPW program,
Connect the scale to a computer/switch using a PT0017 or PT0014
cable (depending on the Ethernet module on board – see ch. 27.1.2),
Set the port for communication with a computer – parameter
<P3.1.1 COMP. PORT> set to Ethernet (see ch. 15.1.1),
- 164 -