User Manual Scales of Y/KTPS series Manual number: ITKU-51-03-01-11-A Control of Packaged Goods Network version • WLY/KTPS Scales • WPY/KTPS Scales MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS RADWAG Wagi Elektroniczne, 26–600 Radom Bracka 28 Street - POLAND Phone +48 48 38 48 800, phone/fax. +48 48 385 00 10 Selling department +48 48 366 80 06 www.radwag.
JANUARY 2011 2
Table OF CONTENTS 1. 2. 3. 4. INTENDED USE..........................................................................................................................5 PRECAUTIONARY MEASURES................................................................................................5 WARRANTY CONDITIONS........................................................................................................6 UNPACKING AND MOUNTING .............................................................................
14.1.2. Computer address..................................................................................................27 14.2. Printer................................................................................................................................27 14.2.1. Printer port .............................................................................................................28 14.2.2. Printer code page ....................................................................................
1. INTENDED USE Scales of WPY/KTPS series together with KTP NET software constitute a multi-stand system (network). Each scale is an independent scale and information about the status of testing are sent in real time. The computer program allows to collect data in real time from every connected scale. The system allows to start test controls from scales or the computer level.
3. WARRANTY CONDITIONS A. RADWAG is obliged to repair or change those elements that appears to be faulty because of production and construction reason, B. Defining defects of unclear origin and outlining methods of elimination can be settled only in participation of a user and the manufacturer representatives, C. RADWAG does not take any responsibility connected with destructions or losses derives from non-authorized or inappropriate (not adequate to manuals) production or service procedures, D.
D. Scale should be levelled by turning regulation feet. Levelling is correct if air bubble is situated in the central position: 4.2. WPY/KTPS/D2 scales A. Take the device out of the package, B. Put the scale on an even stiff ground, C.
E. Scale should be levelled by turning regulation feet. Levelling is correct if air bubble is situated in the central position: 5. SCALE CONSTRUCTION 5.1. Main dimensions 5.1.1. WPY/KTPS/C1 scales 5.1.2.
5.2. Description of connectors 5.2.1. Connectors’ description in PUE 7 1 – Ethernet RJ45 2 – RS232 (COM1) 3 – USB 1 – power supply socket 2 – I/O, RS232 (COM2) 5.2.2.
5.2.3. Description of glands PUE 7P 1 – Supply cord gland 2 – Gland for platforms 1, 2 3 – Gland for platforms 3, 4 5.2.4.
6. GETTING STARTED • After the terminal is connected to power the ON/LOAD diode starts to light. to start the operating system loading procedure. • Press Windows CE together with RADWAG software loading is signalled by blinking the red diode ON/LOAD. • When the loading procedure is completed the main software window appears. 7.
8. FUNCTIONS OF KEYS 8.1. Keys on the overlay Key Description Turning on/off the scale Zeroing Tarring Confirmation of weighing Entering the scale menu (not active after establishing a connection with the computer) Selecting products Selecting operators Batch quantity declaration 8.2.
Moving left the description line for button functions. Scrolling up the list of operators or products Moving right the description line for button functions. Scrolling down the list of operators or products Leaving the screen displaying the list of operators or products 9. INDICATING WINDOW In the main application window one can see four separate parts: 1. The working mode information is displayed in the top part of the display. It is accompanied by date and time.
• If a scale establishes connection with the database following messages are displayed in the workspace: Notice: Description of setting codes and cooperation with PC software KTP NET 2003 can be found in ch. 18 of this manual. 4. There are screen buttons below the workspace (description of button functions – see ch. 8.2 of this manual):: 10. PROGRAM STRUCTURE The main menu has been divided into six functional groups. In every group there are parameters of similar use.
10.1. Inventory of parameters 10.1.1. Scale parameters - weighing Icon Description Value Platform 1 - Median Filter 0.5 Filter Fast Autozero Yes LO threshold 0 Weighing software version 2.0 10.1.2.
Parity None Ethernet - DHCP No IP Address 192.168.0.2 Subnet mask 255.255.255.0 Gateway 192.168.0.1 Tcp - Port 4001 10.1.3.
Code length 0 Transponder card reader - Port None Description Value Actions - Left proximity sensor None Right proximity sensor None Input 1 None Input 2 None Input 3 None Input 4 None Set Default - Show all platforms No 10.1.4. Display Icon 10.1.5.
Touch screen calibration - Factory Number - Software version 2.0.0 KTP NET 10.1.6. User Calibration An option only for non-verified scale Icon Description Value Platform 1 - Setting of start mass - Calibration - 10.2. Navigating within the menu Owing to the colour display with the touch panel navigating within the menu is simple and intuitive. 10.2.1.
Decreasing parameter value Increasing parameter value Enter (OK) Abort Clearing an editing field Screen keyboard on / off One level up 10.2.2. Return to weighing The changes introduced are saved for good after they are several times until the following confirmed. Press message box appears: Press: – to confirm changes or changes. The program returns to weighing.
11. WEIGHING Put a load on the weight pan. When pictogram is ready to read. is displayed the indication 11.1. Conditions of operational use In order to assure a long term operating period with appropriate measurements following principles should be adhered to: • Avoid applying mechanical shocks to the weight pan: • Loads should be placed in the centre of the pan (eccentric errors are outlined in PN-EN 45501 chapter 3.5 and 3.6.
• Avoid side loads, particularly side strokes: 11.2. Zeroing In order to zero the indication choose a platform on the touch panel and . press After zeroing is performed the indication is equal zero and following and . symbols usually appear: Zeroing is possible only when the indication is stable. Notice: Zeroing is possible only within ±2% of full range around zero. If the zeroed value is beyond the interval of ±2%, Err2 is displayed. 11.3.
11.4. Weighing for dual range scales Switching between the I range and the II range happens automatically (exceeding Max of the I range). Weighings in the second range is signalled by a pictogram in the top left . Then weighings is done with the accuracy corner of the display ) of the II range to the moment of returning to zero (autozero range where the scale switches back to the I range. Switching between the II range and I range is automatic both in the switching point the autozero zone.
Inventory of scale parameters: Median Filter Filter Autozero LO Threshold 12.1. Median filter The median filter is intended for eliminating short-lasting mechanical shocks. Procedure: • Enter < < Weighing> according to ch. 12 of the manual, select Median Filter> and then set an appropriate value. Accessible settings: None - median filter is off 0.5, 1, 1.5, 2, 2.5 - filtering level to choose 12.2.
Notice: The higher filtering level the longer stabilization time. 12.3. Autozero The autozero function has been implemented in order to assure precise indications. This function controls and corrects „0” indication. While the function is active it compares the results continuously with constant frequency. If two sequentional results differ less than the declared value of autozero range, so the scale will be automatically zeroed and the and will be displayed.
13. COMMUNICATION The scale can communicate with external devices via different ports: • • COM 1 (RS232), COM 2 (RS232), • Ethernet, • Tcp. The communication can be configured in parameters’ group Communication>. < In order to enter < < Communication>, press and then Communication>. 13.1. RS 232 settings Procedure: • Enter < Communication> according to ch.13 of the manual, select < COM1> or < COM2>, and then set an appropriate value.
Following settings are accessible for Ethernet: • • • • DHCP IP Address Subnet Mask Default gateway - Yes – No - 192.168.0.2 - 255.255.255.0 - 192.168.0.1 Notice: The settings above are only for information purposes. Transmission parameters should be matched to the local client network. • Go back to weighing saving parameters and restart the device. 13.3. TCP protocol setting TCP (Transmission Control Protocol) is a protocol for communication between two computers. It operates in mode client-server.
14.1.1. Computer port Procedure: • Enter parameters’ group < 14 of this manual, select „ set the appropriate option. Devices> according to ch. Computer / Port” and then The scale can communicate with a computer via following ports: • RS 232 (COM1), • RS 232 (COM2), • Tcp. 14.1.2.
14.2.1. Printer port Procedure: • Enter < Devices> parameter group as described in chapter 14 of the manual, choose „ appropriate option. Printer / Port” and then select an Printers can be attached to: • RS 232 (COM1), • RS 232 (COM2), • USB. 14.2.2. Printer code page Procedure: • Enter parameters < of the manual, Printer / • Choose „ will be displayed, Devices> as described in chapter 14 Code Page” then the screen keyboard • Write the required code page and confirm by pressing .
Notice: Communication> set the baud rate (default 9600b/sec). In submenu < The detailed description of cooperation scale – barcode scanner can be found in APPENDIX A in this manual. 14.3.1. Port for barcode scanner Procedure: • Enter < „ Devices> according to ch.14 of the manual, choose Barcode reader / Port” and then set the appropriate value. Barcode scanners can be connected to: • RS 232 (COM1), • RS 232 (COM2), 14.3.2.
Procedure: • Enter < Devices> according to ch.14 of the manual, • Choose „ Barcode Scanner / keyboard is displayed, Code Length” then the screen • Write a new length and confirm it by pressing . 14.4. Transponder card reader Depending on settings in PC software KTP NET, selecting (logging on) an operator, after establishing connection with the computer, can be performed by: • Typing a password on a keyboard, • Approaching a transponder card to the reader. The card needs to be registered first.
15. DISPLAY 15.1. Function keys In submenu < • • Actions> users can set actions following keys: proximity sensors, digital inputs of PUE7. If a sensor or input has been attributed a function it has been activated at the same time. If a sensor or input has no ascribed a function in stays inactive. Procedure: • Press and then: „ Display / Actions”, • Ascribe actions for optical sensors and/or digital inputs.
Assort Logout F1 F2 F3 F4 F5 Arrow up Arrow down G T *) – Functions not attached to software KTP NET. 15.2. Displaying platforms If a terminal is equipped with two platforms users can switch between platforms in two ways: • By pressing the platform number on the scale screen, • By activating in parameters all platforms that will be separately placed in the main window of the program. In that case platforms can be activated by pressing the area of this platform.
To activate all platforms press „ Display / , choose: Show all platforms”, and set appropriately. NO YES - Displaying all platforms disabled Displaying all platforms enabled 16. OTHER PARAMETERS There is a group of parameters different from others which influence the operation of the scale. They are gathered in group < To enter < Others>, press Others> e.g. language, beep etc. and then < Others>. 16.1. Languages Procedure: • Enter submenu < choose < Others> according to ch.
Procedure: • Enter < Others> according to ch. 16 of this manual, • Select option < Date and Time> then the screen keyboard appears, • Set year, month, day, hour, minutes and confirm by pressing . 16.3. Sound signal Procedure: • Enter < Others> according to ch. 16 of this manual, < and set accordingly.
16.5. Software version Procedure: • Enter submenu < choose < number. Others> according to ch. 16 of this manual, Software Version> to see the program version 16.6. Factory number Procedure: • Enter submenu < choose < Others> according to ch. 16 of this manual, Factory number> to see the program version number. 17. CUSTOMER CALIBRATION An option only for non-verified scale Scales require to recalculate internal divisions to more suitable ones (e.g. g, kg etc.).
• Take the load off the pan of platform 1, . The following message appears during • Press button adjusting start mass: „Evaluation of starting mass”, • After the procedure has been completed the following message box appears: • Put calibration mass on pan of platform 1 and then select , • After the procedure of calibration factor determination following command appears: • Return to weighing, saving parameters. Setting of start mass> parameter allows to adjust < start mass of platform 1.
Return to weighing: The changes introduced are saved for good after they are confirmed. several times until the Press following message box appears: – to confirm changes or Press: changes. The program returns to weighing. – to abort 17.2. Start mass adjustment It is possible to adjust only a start mass, it helps to correct the start zero when the span does not change. Procedure: • Enter submenu < and select: “ Customer Calibration> according to ch.
• After the procedure has been completed the following message box appears: • Return to weighing, saving parameters. 18. COOPERATION WITH PROGRAM KTP NET 18.1. Connection with computers In order to establish connection with computer software KTP NET : • In submenu < Computer> choose the communication port and scale address (see ch. 14.1 of this manual), Communication> configure the selected port • In submenu < (see ch. 13 of this manual), • Connect the scale and computer using an appropriate cable.
Notice: Cable „scale - Ethernet” is the standard Ethernet cable terminated by connector RJ45 . After establishing connection of a scale with the database in the PC software, following messaged are displayed in the workspace: Notice: In the last fourth line of the work area there are descriptions of button functions. Using and this line can be moved on the screen. In order to start the test control choose: • • • • Product, Operator, Batch quantity, Batch number. 18.2.
18.2.1. Choosing products by entering identifier Procedure: • Press , then you will see dialog box with the screen keyboard, • Enter the required product identifier and press then software returns to the main window. to confirm it, Notice: If no product with the entered identifier does not exist in the database there is a message box displayed for 2 seconds: . The weighing program returns to displaying . 18.2.2.
• In order to confirm the selected item press • Press to leave the list without changes. , 18.2.3. Product code related searching Procedure: • Press button is displayed, , then the list of products from the database , and then a message and the list of products complemented by codes is displayed, • Using buttons or set marker „>” next to the required product, • When the number of records in the database is significant users can search the database. See ch. 18.2.
18.3.1. Choosing operators by entering identifier Procedure: • Press , then you will see dialog box with the screen keyboard, • Enter the required operator identifier and press then software returns to the main window. to confirm it, Notice: If no operator with the entered identifier does not exist in the database there is a message box displayed for 2 seconds: . The weighing program returns to displaying . 18.3.2.
• When the number of records in the database is significant users can search the database. See ch. 18.2.2., • In order to confirm the selected item press • Press , to leave the list without changes. 18.3.4. Choosing operators using RFID readers In order to log in using an RFID transponder card reader the card need to be attributed to an operator in the database of PC software. The mode of saving transponder cards numbers can be enabled form the level of program KTP NET.
18.6. Starting test control After all the necessary codes are entered following information are displayed in the main window workspace: • • • • • • Operator name, Product code and name, Product nominal mass (Qn), Tare ascribed to the product (for controls with average tare), Batch quantity, Test control mode.
[1] Product test - Product code and name Qn - Product nominal mass TARE - Tare value attributed to the product PUT 1 / 30 - Command concerning putting subsequent products that are numbered out of the whole sample T - Number of errors T (Lower than Qn –T) in the sample 2T - Number of errors 2T (Lower than Qn –2T) in the sample Th. 1 - Sample number av. - Average mass of controlled goods OK - Test results: OK.
Procedure: • After starting a test control following information data are displayed in the display workspace: [1] Product test - Product code and name Qn - Product nominal mass Load empty: 1/10 - Command concerning putting subsequent products that are numbered out of the whole charge 1/30 - Number of performed full measurements out of the sample quantity TARE VALUES COLLECTING - Message for command „Load empty” T - Number of errors T (Lower than Qn –T) in the sample 2T - Number of errors
[1] Product test - Product code and name Qn - Product nominal mass Load full 1/10 - Command concerning putting subsequent products that are numbered out of the whole charge 1/30 - Number of performed full measurements out of the sample quantity GROSS WEIGHINGS COLLECTING T - Message for command „Load full” - Number of errors T (Lower than Qn –T) in the sample 2T - Number of errors 2T (Lower than Qn –2T) in the sample Av.
Deviation - Standard deviation Adm. - Acceptable value of standard deviation Test in progress - Status F5-End - Description of button terminating control NET - Net mass of tested package Notice: In order to have a product tested in mode „Average Tare”, standard deviation „Deviation” of weight of tested packages (sample of at least 10 measurements) cannot exceed 0.25 of the maximum acceptable negative limit T with reference to the nominal package weight.
19. TECHNICAL PARAMETERS 19.1.
19.2.
20.
21. APPENDIX F - Communication with barcode scanners 1. For communication with barcode scanners RADWAG scales use RS232 interfaces and simplex transmission (one direction) without handshaking. Only two wires are required for assuring such a transmission. Used scanners should be equipped in such interface with disabled both hardware and software handshaking. 2. Both scales and scanners have the possibility of setting of transmission parameters.
MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS RADWAG WAGI ELEKTRONICZNE 26 – 600 Radom, Bracka 28 Street POLAND Phone +48 48 38 48 800, phone/fax. + 48 48 385 00 10 Selling Department + 48 48 366 80 06 www.radwag.