After placing the load, scale display will indicate net mass. Tarring can be
done many times in full measuring range of scale.
While using the tare function, pay attention not to exceed max measuring
range of scale. After removing load and packaging indication equal to sum
of tarred masses with minus sign will be displayed.
Notice:
Tarring process can not be carried out when there is negative mass
value or zero mass value on scale display. In such case display
window < WEIGHT kg> will indicate message <Err3>.
13.2. Scale zeroing
In order to zero mass indication press key . In display window
<WEIGHT kg> symbols ▼ will be indicated above ZERO and STA.
Zeroing is the same as setting new zero point, which is treated by the scale
as precise zero. Zeroing is possible only when display state is stable.
Notice:
Zeroing display state is possible only in range up to ±2% of scale max
capacity. If zeroed value will be higher than ±2% of max capacity, display
window <WEIGHT kg> will indicate message <Err2>.
14. MAIN PARAMETERS
User can adjust scale to external environmental conditions (filer level) or
to his own needs (autozero function). These parameters are in group
<P1. rEAd>.These functions will help the user to adjust the scale to
environment al conditions in which scale operates.