User manual
- 38 -
05 Display time
Available settings
1 – YES time displaying enabled, the upper bargraph of the display contains time,
0 – NO time displaying disabled.
06 Display date
Available settings
1 – YES date displaying enabled, the upper bargraph of the display contains date,
0 – NO date displaying disabled.
Return to weighing mode
(see point 6.1.2. – Return to weighing mode)
11. SETTING BALANCE OPERATING PARAMETERS
Balance XA/2X series, in menu group <P4 Readout> enables adjusting balance operation
to current ambient conditions at a workstation (filter) and required user needs (display
refreshment, autozero, previewing last digit).
Fig. 22. Submenu Readout – internal settings
11.1. Filter settings
Depending on the ambient conditions at a workstation, the balance enables setting the
filtering value. In case of very good operating conditions it is recommended to set the
filtering value to very fast (parameter value 01 Filter set to 1). If the operating conditons are
harsh (air drafts, vibrations) set the filter to slow or very slow (parameter value 01 Filter set