Before startup of pipette calibration process set:
- Pipette type [volume: fixed / adjustable]
- Pipette volume Vmax [ml]
- Pipette volume V1/2max [ml]
- Pipette volume Vmin [ml]
- Volume error at Vmax [%]
- Volume error at V1/2max [%]
- Volume error at Vmin [%]
- Liquid temperature [oC]
- Atmospheric pressure [hPa]
- Humidity [%]
- Number of samples
Display content
Fig. 55. Pipette calibrationdisplay content
1 – measurement result
2 working mode name
3 tested volume
4 average volume
5 accuracy error
6 number of samples (measurements)
The procedure is completed by determining the accuracy error [ES], standard deviation [sr]
and repeatability error [CV].
On each checking procedure the graphic display of a balance indicates the checking result.
Fig. 56. Pipette calibration processsummary