User manual

- 93 -
Command SU CR LF sends result in current measuring unit on stabilization of
indication)
Function IMMEDIATELY SEND MEASUREMENT RESULT IN CURRENT
MEASURING UNIT
Command SUI CR LF
Function ZERO BALANCE
Command Z CR LF (zeroing of indication on stabilization)
Function IMMEDIATELY ZERO BALANCE
Command ZI CR LF
Function TARE WHEN STABLE
Command T CR LF
Function IMMEDIATELY TARE BALANCE
Command TI CR LF
Function SWITCH OFF CONTINUOUS TRANSMISSION IN BASIC MEASURING
UNIT
Command C0 CR LF
Function SWITCH ON CONTINUOUS TRANSMISSION IN BASIC MEASURING
UNIT
Command C1 CR LF
Function SWITCH OFF CONTINUOUS TRANSMISSION IN CURRENT
MEASURING UNIT
Command CU0 CR LF
Function SWITCH ON CONTINUOUS TRANSMISSION IN CURRENT MEASURING
UNIT
Command CU1 CR LF
Function GIVE FACTORY NUMBER
Command NB CR LF
Function GIVE MEASURING RANGE
Command FS CR LF
Function GIVE SOFTWARE REVISION
Command RV CR LF
Function GIVE OR CHANGE DATE IN A BALANCE
Command PD CR LF (causes sending data on set date or changes the date)
Function GIVE OR CHANGE TIME IN A BALANCE
Command PD CR LF (causes sending data on set time or changes the time)