TABLE OF CONTENTS 1. 2. 3. 3.1. 4. 4.1. 4.2. 5. 5.1. 5.2. 6. 6.1. RECOMENDATION ....................................................................................................... 4 INTENDED USE ............................................................................................................ 4 DESIGN ......................................................................................................................... 4 Safety analysis 5 OPERATION – THE MANUAL..............................
1. RECOMENDATION Following recommendations of this user manual is essential condition for ensuring safety when using ZRi02 power supplier. The user, as well as every single person performing maintenance operation on the ZRi02 power supplier, should familiarize with this manual. The manual must be available for reference to any person operating it. The manual may be extended with additional warnings and safety information by RADWAG.
3.1. Safety analysis RADWAG as a manufacturer, guarantees intrinsic safety of the device and compliance with PN-EN 60079-0 and PN-EN 60079-11 standard. • • • • • • • Maximal output power (Po) is lower than 0,52W Some outlined soldering points are the subject of detailed monitoring. The mains power transformer with a thermal fuse protecting the primary winding complies with: PN-EN 60079-11 Resistor limiting current Zener barrier limiting the voltage in the intrinsic safety circuit.
HAZARDOUS AREA EX SAFE AREA DEVICE Cable up to 20m 4.2.
Main subassemblies: 1 – Power cord 230V AC terminated with a plug 2 – LED signalling the presence of output voltage 3 – output supply cable (up to 20m) 4 – Power output plug, for connecting to DC IN 5. SAFETY REQUIREMENTS The ZRi02 power supplier is an accompanied device and can be installed only outside a hazardous area. 5.1. Exchanging fuses Fuses can be exchanged only outside the hazardous area after prior disconnecting from mains (230V AC).
• • • • remove the burnt mount a new fuse, characterized by parameters given below Put on the insulation coating on the fuse (pos.
5.2. Cleaning Cleaning may take place only after prior disconnection of the power supplier from mains 230V AD and from the device it is connected to. It should be cleaned only by means of substances and cloths that prevent accumulation of electrostatic charges. It is forbidden to use abrasives or solvents in order to protect the casing against damage. 6. TECHNICAL DATA • • • • • • • • • • • • • • Uo = 8,61 V.