ToxiRAE Pro PID User’s Guide P/N G02-4008-000 Rev.
© Copyright 2016 RAE Systems by Honeywell.
ToxiRAE Pro PID User’s Guide Contents 1. General Information ........................................................................................................ 7 2. Specifications .................................................................................................................. 8 Radio Specifications ........................................................................................................... 9 3. Overview .................................................................
ToxiRAE Pro PID User’s Guide 13.2. Calibration Adapter...................................................................................... 47 13.2.1. Connecting The Calibration Adapter ............................................................ 47 13.2.2. Disconnecting The Calibration Adapter ....................................................... 48 13.3. Zero Calibration ........................................................................................... 48 13.4. Span Calibration................
ToxiRAE Pro PID User’s Guide WARNINGS Read Before Operating This manual must be carefully read by all individuals who have or will have the responsibility of using, maintaining, or servicing this product. The product will perform as designed only if it is used, maintained, and serviced in accordance with the manufacturer’s instructions. CAUTION! Never operate the monitor when the cover is removed. Remove the monitor top cap, rear cover, fan, or battery only in an area known to be non-hazardous.
ToxiRAE Pro PID User’s Guide WARNINGS Use only RAE Systems lithium- ion battery part number G02-3004-000 (3.7V, 1,800mAh). This instrument has not been tested in an explosive gas/air atmosphere having an oxygen concentration greater than 21%. Substitution of components may impair suitability for intrinsic safety. Replace batteries only in non-hazardous locations. STATIC HAZARD: Clean only with a damp cloth to avoid damage from electrostatic discharge.
ToxiRAE Pro PID User’s Guide AVERTISSEMENT Utiliser seulement la batterie Lithium- ion RAE Systems référence G02-3004-000 (3.7V, 1,800mAh). Cet instrument n’a pas été testé dans une atmosphère de gaz/air explosive ayant une concentration d’oxygène plus élevée que 21%. La substitution de composants peut compromettre la sécurité intrinsèque. Ne changer les batteries que dans un emplacement désigné non dangereux. RISQUE D'ORIGINE ELECTROSTATIQUE: Nettoyer uniquement avec un chiffon humide.
ToxiRAE Pro PID User’s Guide CAUTION This device complies with Part 15 of the FCC Rules / Industry Canada license-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence.
ToxiRAE Pro PID User’s Guide 1. General Information ToxiRAE Pro PID is a personal wireless monitor for volatile organic compounds (VOCs), the world’s smallest. Featuring RAE Systems’ next-generation PID sensor, the ToxiRAE Pro PID can promptly detect and accurately monitor over 300 VOCs. With an onboard library of 190 correction factors, the ToxiRAE Pro PID can be programmed to automatically read in concentrations of the specified compound. Key Features Wireless1 . Personal.
ToxiRAE Pro PID User’s Guide 2. Specifications ToxiRAE Pro PID Specifications Dimensions Weight Battery Charging Time Operating Time Charger 118 mm x 60 mm x 30 mm (4.6" x 2.4" x 1.2") 235g (8.
ToxiRAE Pro PID User’s Guide Radio Specifications Radio model: RM900 Frequency range: Within 902 to 907.5 MHz and 915 to 928 MHz, use IEEE 802.15.4 channel 1, 6, 7, 8, 9 and 10 Modulation: 802.15.4 DSSS BPSK RF power (Tx):10dBm Data rate: 40kbps Wireless Approval For UAE In Middle East TRA REGISTERED No: ER36635/15 DEALER No: HONEYWELL INTERNATIONAL MIDDLE EAST - LTD - DUBAI BR Wireless Approval For QATAR In Middle East ictQATAR Type Approval Reg. No.
ToxiRAE Pro PID User’s Guide 3. Overview The ToxiRAE Pro PID is a wireless, personal volatile organic compounds (VOC) monitor, the world’s smallest. The ToxiRAE Pro PID takes worker protection to the next level by providing safety professionals wireless access to real-time instrument readings and alarm status from any location for better visibility and faster response.
ToxiRAE Pro PID User’s Guide 4.1. Display Overview The LCD display provides visual feedback that includes the sensor types, readings, time, battery condition, and other functions. Calibration and bump test up to date Sensor name and “Alarm” are displayed alternately when in alarm Wireless on/off status Wireless signal strength Datalog on Bump test due Battery status Sensor calibration due Man Down alarm on Reading Unit of measure Alarm type (when in alarm) Soft keys (functions change by activity) 4.
ToxiRAE Pro PID User’s Guide Icon Function Bump test due All sensors tested and calibrated tick mark (all sensors have been bump tested and calibrated; no sensor is overdue for a bump test or calibration according to the intervals configured on the instrument 5. Charging The ToxiRAE Pro PID Always fully charge the battery before using the ToxiRAE Pro PID. The ToxiRAE Pro PID’s Li-ion battery is charged by placing the ToxiRAE Pro PID in its cradle.
ToxiRAE Pro PID User’s Guide Follow this procedure to charge the ToxiRAE Pro PID (the procedure is applicable to the instrument with or without the rubber boot installed): 1. Plug the AC/DC adapter into the ToxiRAE Pro PID’s cradle. 2. Plug the AC/DC adapter into the wall outlet. 3. Place the ToxiRAE Pro PID into the cradle and press down until it is locked in place. The ToxiRAE Pro PID begins charging automatically. The LED in the cradle should glow red to indicate charging.
ToxiRAE Pro PID User’s Guide 6. Turning The ToxiRAE Pro PID On To turn the ToxiRAE Pro PID on, press and hold [MODE] for 3 seconds. Caution: The alarm buzzer is very loud. During startup, you can mute most of the sound by holding a finger over the alarm buzzer port. Note: Never put tape over the alarm buzzer port to permanently mute it or otherwise limit the loudness of the buzzer. 1. When the ToxiRAE Pro PID turns on, the buzzer, vibration alarm, and LEDs are tested.
ToxiRAE Pro PID User’s Guide 4. After self-testing, two start-up options are available for the ToxiRAE Pro PID: Normal and Fast Start-up.
ToxiRAE Pro PID User’s Guide 9. Low Battery Warnings If, during operation, the battery’s charge is low, the battery icon in the display screen turns empty and blinks once every second. At the same time, the following notifications occur every minute: the buzzer beeps and LEDs blink once, followed by a vibration and a screen message “Battery low! Shut off in X minutes.” The countdown to shutoff starts 10 minutes prior to automatic shutoff. 10.
ToxiRAE Pro PID User’s Guide 11. Operating The ToxiRAE Pro PID 11.1. Overview The ToxiRAE Pro PID has two operational modes: • • Normal — see page 17 for detailed instructions. Diagnostic — see page 51 for detailed instructions. The following is an overview of the two modes: Normal Mode is the default mode. It is accessed when you turn on the ToxiRAE Pro PID. Two sub-modes are available: Basic User and Advanced User.
ToxiRAE Pro PID User’s Guide 11.1.2. Basic User Mode When the ToxiRAE Pro PID is in Basic User Mode, pressing [MODE] allows you to step through the screens in the following fashion: Note: The gray boxes indicate the datalog functions when datalogging is in Manual mode. When datalogging is in Automatic mode, these screens do not appear. Refer to page 30 for information on selecting Automatic or Manual datalogging.
ToxiRAE Pro PID User’s Guide 11.1.3. Advanced User Mode Note: To enter Advanced User mode from Basic User mode, you must enter Programming mode on an instrument that is running in Basic User mode by simultaneously pressing and holding both the buttons for 3 seconds and supply a password. To exit Programming Menu and return to the measurement screen (in Advanced User mode), press the [MODE] button repeatedly to highlight the Exit icon in Programming Menu and press Y to exit to the measurement screen.
ToxiRAE Pro PID User’s Guide 11.2.1. TWA, STEL, Peak The first screen includes a readout for TWA, STEL, and Peak readings. TWA The TWA (time-weighted average) reading is the average reading of the gas concentration times that the monitor has been on, prorated to an 8-hour period. STEL Short Term Exposure Limit (STEL) data based on 15-minute STEL values, expressed in ppm or mg/m3 . A STEL value is only displayed after the ToxiRAE Pro PID has been on for 15 minutes or longer.
ToxiRAE Pro PID User’s Guide 11.2.5. Enter Communications and stop measurement? Before communicating with a PC running ProRAE Studio II Instrument Configuration and Data Management software (for downloading datalogs, configuring the instrument, or uploading new firmware), the ToxiRAE Pro PID must be set to communicate with the computer. Press [MODE] repeatedly until you reach the “Enter Communications and stop measurement?” screen. Then press [Y/+].
ToxiRAE Pro PID User’s Guide When you have successfully entered Programming Mode, the following screen appears: Note: The password can only be changed by connecting the instrument to a PC running ProRAE Studio II software. Follow the instructions in ProRAE Studio II to change it. The Calibration icon is highlighted first by default, but you can press [MODE] to step from one programming menu item to the next, with the name of the menu shown at the top of the display and the corresponding icon highlighted.
ToxiRAE Pro PID User’s Guide 11.4.1. Navigating Programming Menus Navigating through the Programming menus is easy and consistent, using a single interface format of a “forward arrow” and “Select,” controlled by the [MODE] key and [Y/+] key, respectively. In the submenus, the functions of a “down arrow” and “Select” are controlled by the [MODE] key and [Y/+] key, respectively. Also in submenus, the functions of the keys may correspond to the instruction directly above each key.
ToxiRAE Pro PID User’s Guide Once you enter Programming, the LCD displays the first menu, Calibration. Each subsequent menu is accessed by pressing [MODE] repeatedly until the desired menu is displayed. To enter a submenu of a menu, press [Y/+]. 11.5.1. Calibration Two types of calibration are available: Zero (fresh air) and Span. In the Calibration submenu, select Zero or Span Calibration by pressing [MODE]. Once your choice is highlighted, press [Y/+].
ToxiRAE Pro PID User’s Guide Span Calib The procedure for performing a basic span calibration is covered on page 49. Exit Press [Y/+] to exit. Calibration is exited, and Measurement is highlighted in the Programming Menu screen. 11.5.2. Measurement The submenus for Measurement are Measurement Gas and Measurement Unit. Meas Gas Measurement gases are organized in a list called Gas Library: The Gas Library consists of 190 compounds found in RAE Systems’ Technical Note TN106 (available online at www.
ToxiRAE Pro PID User’s Guide 11.5.3. Alarm Setting High Alarm Set the High Alarm as follows: 1. In the High Alarm submenu, press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to 9. Once 9 is reached, pressing [Y/+] again “wraps” around back to 0. 2. Press [MODE] to move to the next digit. 3. When you are satisfied with the number, press [MODE] until “OK” is highlighted. 4. To save the new value, press [Y/+]. The value is saved, and the menu advances to Low Alarm.
ToxiRAE Pro PID User’s Guide 4. To save the new value, press [Y/+]. The value is saved, and the menu advances to TWA Alarm. To exit without saving the new value: 1. Press [MODE] to advance to Cancel. 2. Press [Y/+] to exit the STEL Alarm submenu and advance to the TWA Alarm submenu. TWA Set the TWA Alarm as follows: 1. In the TWA Alarm submenu, press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to 9. Once 9 is reached, pressing [Y/+] again “wraps” around back to 0. 2.
ToxiRAE Pro PID User’s Guide Buzzer & Light The buzzer, light, and vibration alarms can be programmed so that any combination, all, or none of them are activated when the ToxiRAE Pro PID goes into alarm.
ToxiRAE Pro PID User’s Guide When the Man Down feature is on and there is no gas alarm, the ToxiRAE Pro PID senses that it is motionless for the amount of time set in the “Motionless Time” parameter. If the instrument is not moved during that time, then a pre-alarm is activated to alert the user, and shows the “Are You OK?” screen. Pressing [Y/+] clears the alarm and returns the ToxiRAE Pro PID to its normal operation.
ToxiRAE Pro PID User’s Guide Exit (the Alarm Settings Submenu) Press [Y/+] to exit the Alarm Settings submenu. Alarm Settings is exited, and Datalog is highlighted. If you want to cycle through the Alarm Setting submenus, press [MODE] instead of [Y/+]. 11.5.4. Datalog Clear Datalog This operation erases all data stored in the datalog. Note: Once the datalog is cleared, the data cannot be recovered. Press [Y/+] to clear the datalog.
ToxiRAE Pro PID User’s Guide Data Selection Data Selection allows you to select which types of data are recorded and made available when you download your datalog to a computer via ProRAE Studio II software. You can choose any or all of four types of data (you must choose at least one): • • • • Minimum Average Maximum Real Time 1. Press [MODE] to step from one option to the next. The highlighter indicates your choice. 2.
ToxiRAE Pro PID User’s Guide 11.5.5. Monitor Setup Wireless There are seven Wireless parameters: • • • • • • • • Radio On/Off Roaming PAN ID Channel Join Network Interval Off Network Alarm Factory Reset Radio On/Off This turns the wireless modem on or off. Note: Turning the radio off does not change any of the settings. They are stored and ready for use the next time the wireless modem is turned on. 1. Choose an “On” or “Off” by pressing [MODE] until your choice is highlighted.
ToxiRAE Pro PID User’s Guide The message “Applying Settings” is shown and then the menu advances automatically to “PAN ID.” PAN ID All units on a network must be programmed with the same Pan ID (Personal Network Identifier) to ensure communication compatibility. Important! The allowed range for Pan ID numbers is 001 through 999. Note: When Roaming is on, you cannot change the instrument’s PAN ID. When Roaming is on, the “PAN ID” is not shown in the Wireless menu. Choose and enter a 3-digit PAN ID: 1.
ToxiRAE Pro PID User’s Guide Join Network This screen is for joining an existing network. When you see this screen, press [Y/+] to join, or [MODE] to exit without joining. While searching for the network, the message “Joining network…” is displayed. If the network is found and successfully joined, then “Join network successful!” is displayed. Press [MODE] to acknowledge the connection and return to the Wireless menu.
ToxiRAE Pro PID User’s Guide Factory Reset CAUTION! Selecting this resets the wireless settings to their original factory configurations. A factory reset of the wireless settings deletes any customized settings and cannot be undone. 1. Press [Y/+] to select “Factory Reset.” You will see the message “Reset wireless settings?” 2. Press [Y/+] for “Yes” or [MODE] for “Exit.” If you select “Exit,” the ToxiRAE Pro exits without performing a factory reset. Exit (the Wireless Submenu) Press [Y/+] to exit.
ToxiRAE Pro PID User’s Guide User Mode The ToxiRAE Pro PID has two user modes: Basic Basic users can only see and use a basic set of functions. Advanced Advanced users can see all screens and take advantage of all available functions. Note: The default value for User Mode is Basic. To change the User Mode: 1. Press [MODE] to step from one option to the other. The highlighting changes each time you press [MODE]. 2.
ToxiRAE Pro PID User’s Guide 2. Press [MODE] to advance to the next digit. Repeat this process until all six digits of the new time are entered. 3. Press [MODE] repeatedly until “OK” is highlighted. 4. Press [Y/+] to save your selection, exit the Time submenu and advance to Temp. Unit. Note: If you do not want to save the change, press [MODE] until “Cancel” is highlighted. Press [Y/+] to exit to Temp. Unit without saving the changes. Temp.
ToxiRAE Pro PID User’s Guide 4. Press [Y/+] to save your selection, exit the Zero At Start submenu, and advance to LCD Contrast. LCD Contrast The display’s contrast can be increased or decreased from its default setting. You may not need to ever change the default setting, but sometimes you can optimize the display to suit extreme temperature and ambient brightness/darkness conditions. 1. Press and hold [Y/+] to select a desired level of contrast. The range is 1 to 100.
ToxiRAE Pro PID User’s Guide 12. Policy Enforcement The ToxiRAE Pro PID can be configured to enforce a facility/company’s requirements that calibration and/or bump testing be performed at specified intervals, and to warn the user that calibration/bump testing is required. In addition, it can require calibration or bump testing and not allow normal operation of the instrument unless calibration or bump testing is performed.
ToxiRAE Pro PID User’s Guide 7. Click on the icon to highlight it, and then click “Select.
ToxiRAE Pro PID User’s Guide 8. In ProRAE Studio II, the AutoRAE 2 Cradle is shown, including its Serial Number, under “Online”: 9. Expand the view to show the ToxiRAE Pro PID in the AutoRAE 2 Cradle by clicking the “+” to the left of the image of the AutoRAE 2 Cradle: 10. Double-click on the icon representing the ToxiRAE Pro PID. 11. Click “Setup.
ToxiRAE Pro PID User’s Guide 12. In the menu that now appears on the left side, click “Policy Enforcement.” It is highlighted, and the Policy Enforcement pane is shown: For “Must Calibrate” and “Must Bump,” you have the options of no enforcement or enforcement (including “Can’t Bypass,” and “Can Bypass”). Must Calibrate. The user is prompted to calibrate the instrument when calibration is due (as set by the calibration interval). There are two programmable options: • • Can’t Bypass.
ToxiRAE Pro PID User’s Guide These are the screens that are shown on a ToxiRAE Pro PID after startup if “Can Bypass” is selected: If “Can’t Bypass” is selected, the display looks like this, and only allows the options of performing the test or shutting down: 12.1.2. ToxiRAE Pro PID Desktop Cradle Or Travel Charger Make sure the AC adapter is connected and that a USB cable is connected to a computer running ProRAE Studio II. 1. 2. 3. 4. 5. Turn on the ToxiRAE Pro PID.
ToxiRAE Pro PID User’s Guide The Policy Enforcement pane is shown: You can select “Must Calibrate” and/or “Must Bump” and then set whether the user must perform the selected operation in order to use the instrument. 15. Once you have made your selections in ProRAE Studio II, you must upload the changes to the instrument. Click the icon labeled “Upload all settings to the instrument.” 16. A confirmation screen is shown. Click “Yes” to perform the upload, or “No” to abort.
ToxiRAE Pro PID User’s Guide 17. Exit ProRAE Studio II. 18. Press [Y/+] on the ToxiRAE Pro PID to exit Communication Mode. 12.2. Deactivating Policy Enforcement 12.2.1. AutoRAE 2 Cradle To deactivate Policy Enforcement when using an AutoRAE 2 Cradle, follow the procedure for changing settings. See page 39 for details. 12.2.2.
ToxiRAE Pro PID User’s Guide 17. When you see this confirmation. Click “Yes.” Uploading will take a few seconds, and this progress bar is shown: 18. When the upload is done, exit ProRAE Studio II. 19. Press [Y/+] on the ToxiRAE Pro PID to exit Communication Mode.
ToxiRAE Pro PID User’s Guide 13. Bump Testing & Calibrating The ToxiRAE Pro PID Note: Calibration intervals and bump test procedures may vary due to national legislation. 13.1. Bump Testing & Calibrating With An AutoRAE 2™ Bump testing and calibrating a ToxiRAE Pro PID can be done with an AutoRAE 2 Automatic Test And Calibration Station. Follow all instructions in the AutoRAE 2 User’s Guide for successful bump testing or calibration. 13.1.1.
ToxiRAE Pro PID User’s Guide Caution! After calibration is complete, remove the Calibration Adapter. When monitoring, never operate the ToxiRAE Pro PID with the Calibration Adapter attached. If the Calibration Adapter is attached during normal operation, inconsistent and lower-thannormal readings will occur because of decreased concentration of the gas being monitored. 13.2.2.
ToxiRAE Pro PID User’s Guide If you make a mistake, you can cycle through the digits by pressing [MODE] and then using [Y/+] to change the number in each position. Note: The default password is 0000. If you do not know the password, do not enter anything. Just press [MODE] repeatedly to highlight OK, and then press [Y/+]. Note: If the ToxiRAE Pro PID is in Programming Mode and you do not press a key within 60 seconds, it exits Programming Mode and re-enters Normal Mode. 1.
ToxiRAE Pro PID User’s Guide • Press [Y/+] if you want to change it. If you entered the Calibration Menu without supplying the correct password, or supplied the correct password but chose not to change the Cal.Gas, you will see a message like this: Cal.Gas: Isobutylene Span=100ppm Apply gas… 2. Press [Y/+] to start calibration. 3. Press [MODE] if you want to quit without calibrating. During calibration, the calibration gas name, its concentration, and the countdown are displayed.
ToxiRAE Pro PID User’s Guide 14. Diagnostic Mode In Diagnostic Mode, the ToxiRAE Pro PID provides raw counts for sensor, battery, and other readings. Most of these screens are useful only to service technicians. Therefore, if you require information on Diagnostic Mode, refer to the ToxiRAE Pro PID Service Guide. The ToxiRAE Pro PID’s Diagnostic Mode can only be accessed at startup time. In Diagnostic Mode, ToxiRAE Pro PID displays readings in raw counts instead of units such as parts per million (ppm).
ToxiRAE Pro PID User’s Guide 15.
ToxiRAE Pro PID User’s Guide 16. Downloading Datalogs And Performing PC-Based Instrument Configuration and Firmware Upgrades The ToxiRAE Pro PID communicates with a PC running ProRAE Studio II Instrument Configuration and Data Management software (version 1.03 or higher) to download datalogs, configure the instrument, or upload new firmware. The ToxiRAE Pro PID must be connected to a PC through the supplied Charging / PC Communications Cradle and must be in the PC communications mode. 1.
ToxiRAE Pro PID User’s Guide 17. Maintenance WARNING! Maintenance should be performed only by a qualified person who has proper training and fully understands the contents of this manual. The following guidelines should be followed when changing the battery: 1. Turn off the unit before changing the battery. 2. When replacing the battery, take note of its orientation. The connectors on the circuit board should go into the matching slots in the battery.
ToxiRAE Pro PID User’s Guide 4. Remove the four screws on the back of the instrument by turning them counterclockwise. 5. Hold the ToxiRAE Pro PID down and separate the rear cover from the front by pulling from the top and tilting it open without disconnecting the ribbon cable connecting the two halves. Ribbon cable 6. Carefully disconnect the battery from the battery connectors on the circuit board by pulling it toward the bottom of the instrument. Be careful not to damage the ribbon cable.
ToxiRAE Pro PID User’s Guide 7. Before installing a new battery, visually inspect the contact pins on the ToxiRAE Pro PID circuit board to make sure they are not bent. If they are, carefully straighten them before installing the battery. Install the new battery (part number G02-3004-000), making sure it is inserted fully. Make sure the battery connectors on the circuit board go all the way into the connector slots in the battery. 8.
ToxiRAE Pro PID User’s Guide 11. Replace the top cap and tighten the Philips screw. Important! Charge the battery completely before using the ToxiRAE Pro PID, and check the calibration before placing the instrument in service. 17.2. Proper Battery Disposal The Waste Electrical and Electronic Equipment (WEEE) directive (2002/96/EC) is intended to promote recycling of electrical and electronic equipment and their components at end of life.
ToxiRAE Pro PID User’s Guide 17.3. Removing And Installing The Rubber Boot The protective rubber boot on the ToxiRAE Pro PID must be removed before the sensor top cap or back of the instrument can be removed. When removing the boot, follow these steps outlined here. 1. Flip up the tab on the belt clip on the rear of the instrument. 2. Remove the Philips screw that holds the clip in place. 3. Stretch the bottom corners of the boot and gently slide the bottom of the boot over the rear of the ToxiRAE Pro PID.
ToxiRAE Pro PID User’s Guide 17.4. Filter Replacement A filter is permanently attached inside the ToxiRAE Pro’s top cap, which helps to keep debris and moisture out. This filter should be inspected periodically, and the cap and rubber insert should be inspected for damage. If the filter is dirty or the cap or rubber insert shows signs of damage, replace the entire instrument cap assembly, including the cap, screw, filter, and the rubber insert (part number G02-0305-003, package of three).
ToxiRAE Pro PID User’s Guide 3. Once the top cap is removed, install the new top cap assembly. 4. Place the new rubber insert onto the ToxiRAE Pro PID, being careful to align the insert correctly and that it is not distorted. Also, make sure that you do not press on the fan. 5. Press the new top cap onto the ToxiRAE Pro until it is fully seated. Check to see that the pin on the front side of the cap went into the corresponding groove on the front side of the instrument housing.
ToxiRAE Pro PID User’s Guide 17.5. Cleaning Or Replacing The PID Follow the instructions for removing the top cap in the previous section. 1. Gently lift out the PID module with your fingers, being careful not to damage the fan: Fan Sensor module 2. If the module requires replacement (because the lamp does not illuminate, or the module has been used past its expiration date), place a new module into the slot, being careful to match the indexing keys. The sensor can only go into its slot one way.
ToxiRAE Pro PID User’s Guide 5. Once the cap is removed, set it aside. 6. Now lift the sensor electrode panel from the module: 7. Clean the sensor electrode panel in a solution of methanol lamp cleaner (included, along with cleaning swabs, in a PID Lamp Cleaning Kit, part number 500-0014010, package of 10), and allow it to dry. 8. Clean the lamp’s window with a cleaning swab dipped in methanol lamp cleaner, and allow it to dry.
ToxiRAE Pro PID User’s Guide 10. Reassemble the sensor module by placing the sensor electrode panel back in place and firmly pressing the cap back onto the top. 11. Place the sensor module back into the ToxiRAE Pro PID. Make sure the index points are aligned (it can only go in one way). Match index points on sensor module with ToxiRAE Pro PID 12. Carefully place the cap back onto the ToxiRAE Pro PID and tighten the screw. Note: Always calibrate the ToxiRAE Pro PID after replacing the sensor module. 17.6.
ToxiRAE Pro PID User’s Guide 3. Inspect the new fan and make sure its two electrical pins are straight. 4. Align the fan so that the indexing key (the raised rectangular portion) on its body matches with the keyway in the slot and the electrical pins and carefully slide into the contact points at the bottom of the slot.
ToxiRAE Pro PID User’s Guide 17.7. Cleaning The Exterior Occasional external cleaning of the instrument with a soft cloth is recommended. Do not use detergents or chemicals. If necessary, you can use a damp cloth (water only). It is a good idea to install the Calibration Adapter before cleaning the ToxiRAE Pro PID’s housing, to keep dirt, dust, or moisture away from the sensor openings and to keep the filter clean.
ToxiRAE Pro PID User’s Guide 18. Troubleshooting Problem Cannot turn on power Lost password Reading abnormally Low Possible Reasons & Solutions Reasons: Depleted battery. Defective battery. Solutions: Solutions: Reasons: Solutions: Buzzer, LEDs, or vibration alarm Inoperative Reasons: Solutions: 66 Replace battery. Use ProRAE Studio II software to change the password. Call Technical Support at +1 408-752-0723 or toll-free at +1 888-723-4800 Calibration Adapter is attached. Incorrect calibration.
ToxiRAE Pro PID User’s Guide 19. Technical Support To contact RAE Systems Technical Support: Monday through Friday, 7:00AM to 5:00PM Pacific (US) Time Phone (toll-free): +1 877-723-2878 Phone: +1 408-952-8200 Fax: +1 408-952-8480 Email: RAE-tech@honeywell.com Outside the Americas: E-Mail: HAexpert@honeywell.com Honeywell Analytics Ltd.
ToxiRAE Pro PID User’s Guide 21. Appendix A: Regulatory Information SAFETY INSTRUCTIONS Read Before Operating This manual must be carefully read by all individuals who have or will have the responsibility of using, maintaining, or servicing this product. The product will perform as designed only if it is used, maintained, and serviced in accordance with the manufacturer’s instructions. The user should understand how to set the correct parameters and interpret the obtained results.
ToxiRAE Pro PID User’s Guide Operation Area and Conditions Hazardous Areas classified by Zones ToxiRAE Pro PID (PGM1800) are intended to be used in hazardous areas classified zone 0, zone 1 or zone 2, where gases of explosion groups IIA, IIB or IIC and T4 may be present. For North America and Canada the equipment can further be used in Class I, Zone 0. Hazardous Areas classified by Divisions ToxiRAE Pro is intended to be used in hazardous areas classified for Class I Div.
ToxiRAE Pro PID User’s Guide Year of manufacture To identify the year of manufacture, refer to the serial number of the instrument. The second to last digit in the serial number indicates the year of manufacture. For example, “M” indicates the manufacturing year is 2010.
RAE Systems by Honeywell World Headquarters 3775 N. First St. San Jose, CA 95134-1708 USA Phone: 408.952.8200 Fax: 408.952.8480 E-mail: customerserv@raesystems.com Web Site: www.raesystems.coms P/N G02-4008-000 Rev.