User's Manual

10
(6) Warning
Environment of Use and Storage
Follow the specified process of operational instructions written in
this manual for the safety of the users and patients.
Does not use or store the instrument near any flammable
chemicals such as thinner, benzene, etc. Also, this instrument is
not a category AP or APG equipment. If chemicals are spilled or
evaporate, it may result in fire or electric shock through contact
with electric parts inside the instruments. Also, some
disinfectants are flammable. Be sure to take care when using
them.
Connection
Do not connect the instrument with anything other than specified.
Otherwise, it may result in fire or electric shock.
To avoid the risk of electric shock, this equipment must only be
connected to a supply mains with protective earth.
Wireless connection
SSID & PSK value should match to Routers setting. (Refer to the
Ch.4.(2)). If these values are not matched with Detector and
Router, the connection is not allowed for security.
Avoid using co-channel or adjacent-channel with other wireless
devices for high quality wireless performance.
Handling
Always be sure to keep checking the condition of the system and
the patient to ensure they are normal during the use of the
instrument. If any problem is found, take appropriate measures,
such as stopping the operation of the instrument, as required.
Make sure
to observe
the following right