User's Manual

69
7. Operation
(1) Recommend
X-ray detector should be used at stable state within driving
temperature range.
Acquire the X-ray images after power on and 5 minutes warming
up to obtain high quality images.
The calibration should be performed every 6 months
(2) Switching power on / off
All connection should be done, before turn on the power.
Press the power button by more than 3 sec, when power
on/off.
The green light of the LED indicator on the detector is on, the
detector power is on.
The blue light of the LED indicator on the detector is blinking,
the detector is getting prepared to work and initialize.
After power off, separate the battery.
(3) Storage
Store the sensor unit in clean and dry place. Ensure that storage
place should be not affected by dust or liquids.