User's Manual

58
R-USM-503
ethanol, please also consult a specialist, because they may harm
the CFRP plate.
(3) Inspection
In order to ensure that the instrument is used safely and normally,
please be sure to inspect the instrument before use. If any
problem is found during the inspection, please take measures
indicated in this chapter. If problem still cannot be corrected,
please contact Rayence representative or distributor. It is
recommended that a record of the inspection be kept by making
copies of the check lists in this section, or making a separate
check list.
Inspection chart
Inspection
User
Vendor
Cycle
Check that cables are not
damaged or cover of cables is
not torn
O
Daily
Check that the plugs and locks
of connectors are not loose
O
Daily
Check that the cover or parts
are not damaged and not loose
O
Daily
Check the LED indicator
O
Daily
Check the bad pixel map
O
Half year
Check the performance of the
instrument by performing
exposures using a phantom or a
resolution chart
O
Yearly
For
safety
reasons, be
sure to turn OFF the
power of the
detector when the
following
inspections are
going to be
performed.
Otherwise, it may
result in electric
shock.