REGISTER NAME LOW LIMIT HIGH LIMIT
FACTORY
SETTING
ACCESS COMMENTS
FREQUENTLY USED REGISTERS
40001 0 Input Relative Value (Hi word)
-199999 999999 N/A Read Only
Process value of present input level. This value is affected
by Input Type, Resolution, Scaling, & Tare (Offset) Value.
(Relative Value = Gross (Absolute) Input Value - Tare Value)
40002 1 Input Relative Value (Lo word)
40003 2 Maximum Value (Hi word)
-199999 999999 N/A Read/Write
Maximum Relative Input Capture Value obtained since
having been reset.
40004 3 Maximum Value (Lo word)
40005 4 Minimum Value (Hi word)
-199999 999999 N/A Read/Write
Minimum Relative Input Capture Value obtained since
having been reset.
40006 5 Minimum Value (Lo word)
40007 6 Total Value (Hi word)
-199999 999999 0 Read/Write Totalizer value
40008 7 Total Value (Lo word)
40009 8 Setpoint 1 Value (Hi word)
-199999 999999 100 Read/Write Active List (A or B)
40010 9 Setpoint 1 Value (Lo word)
40011 10 Setpoint 2 Value (Hi word)
-199999 999999 200 Read/Write Active List (A or B)
40012 11 Setpoint 2 Value (Lo word)
40013 12 Setpoint 3 Value (Hi word)
-199999 999999 300 Read/Write Active List (A or B)
40014 13 Setpoint 3 Value (Lo word)
40015 14 Setpoint 4 Value (Hi word)
-199999 999999 400 Read/Write Active List (A or B)
40016 15 Setpoint 4 Value (Lo word)
40017 16
Setpoint 1 Band/Dev. Value
(Hi word)
-199999 999999 0 Read/Write
Active List (A or B).
Applicable only for Band or Deviation Setpoint Action.
40018 17
Setpoint 1 Band/Dev. Value
(Lo word)
40019 18
Setpoint 2 Band/Dev. Value
(Hi word)
-199999 999999 0 Read/Write
Active List (A or B).
Applicable only for Band or Deviation Setpoint Action.
40020 19
Setpoint 2 Band/Dev. Value
(Lo word)
40021 20
Setpoint 3 Band/Dev. Value
(Hi word)
-199999 999999 0 Read/Write
Active List (A or B).
Applicable only for Band or Deviation Setpoint Action.
40022 21
Setpoint 3 Band/Dev. Value
(Lo word)
40023 22
Setpoint 4 Band/Dev. Value
(Hi word)
-199999 999999 0 Read/Write
Active List (A or B).
Applicable only for Band or Deviation Setpoint Action.
40024 23
Setpoint 4 Band/Dev. Value
(Lo word)
40025 24 Setpoint Output Register (SOR) 0 15 0 Read/Write
Status of Setpoint Outputs. Bit State: 0 = Off, 1 = On.
Bit 3 = SP1, Bit 2 = SP2, Bit 1 = SP3, Bit 0 = SP4.
Outputs can only be activated/reset with this register when
the respective bits in the Manual Mode Register (MMR) are
set.
40026 25 Manual Mode Register (MMR) 0 31 0 Read/Write
Bit State: 0 = Auto Mode, 1 = Manual Mode
Bit 4 = SP1, Bit 3 = SP2, Bit 2 = SP3, Bit 1 = SP4,
Bit 0 = Linear Output
40027 26 Reset Output Register 0 15 0 Read/Write
Bit State: 1 = Reset Output, bit is returned to zero following
reset processing; Bit 3 = SP1, Bit 2 = SP2, Bit 1 = SP3,
Bit 0 = SP4
40028 27 Analog Output Register (AOR) 0 4095 0 Read/Write
Functional only if Linear Output is in Manual Mode.
(MMR bit 0 = 1)
Linear Output Card written to only if Linear Out (MMR bit 0)
is set.
40029 28
Input Gross (Absolute) Value
(Hi word)
-199999 999999 N/A Read Only
Gross (absolute) value of present Input level. This value is
affected by Input Type, Resolution, Scaling, but not affected
by Offset Value
40030 29
Input Gross (Absolute) Value
(Lo word)
40031 30 Tare Value (Hi word)
-199999 999999 0 Read/Write
Relative Input Value (standard meter value) is the difference
between the Gross (absolute) input value and the Tare
value, i.e. Relative = Gross - Tare
40032 31 Tare Value (Lo word)
PAX2S MODBUS REGISTER TABLE
Only frequently used registers are shown below. The entire Modbus Register Table can be found at www.redlion.net.
Values less than 65,535 will be in (LO word). Values greater than 65,535 will continue into (Hi word). Negative values are represented by two’s complement of the
combined (Hi word) and (LO word).
Note 1: The PAX2S should not be powered down while parameters are being changed. Doing so may corrupt the non-volatile memory resulting in checksum errors.