Instruction manual

Mounting & Installation
Page
12
Chapter
04
Vögtlin Manual
red-y compact series
Electrical Connection / Power Supply
With the battery-operated devices, mount the battery module. Then the measuring device is ready
for use. The useful life of the battery module is about 2 years, but this depends on use. With
constantly changing flow rates, the useful life may be shorter. We advise you to use only original
battery modules.
With the external power supply and alarm options, you receive a connector cable with loose ends.
The assignment is printed on the cable. You have the option of turning the cable outlet by 180º. To
do this, you open the cover in back by hand and turn the plastic block to the desired direction.
Turning the cable outlet by 180°
The supply voltage must be + 24 Vdc (+/-10%) and have the lowest possible residual ripple. Please
note that voltage drops may occur with longer connecting cables.
Likewise, be aware of possible earth loops when you ground electrically conductive pipes.
Gas Supply
We recommend that you pay close attention to the gas supply line. Contamination in the form of
water, oil or dust are harmful to any measurement principle. Particularly in the case of air supply
with compressor systems, cleanliness cannot always be guaranteed. In case of doubt, install the
appropriate filter. If use-related backflows are to be expected, this filter should also be installed at
the outlet. Please monitor possible pressure loss due to the filter elements.
The gas supply should be oversized and have the capacity of at least twice the flow rate of the
attached measuring device. Monitor the performance of the pressure-reducing unit, as well. Never
install the pressure regulator directly in front of the measuring device.
Check for possible pressure loss along the pipeline system. Flow regulators especially require
defined pressure ratios for proper functioning.
With very low flow rates, pressure reduction is oversized and gas is only periodically fed (on-off
regulation). This becomes evident in the form of periodic fluctuations in the regulation or changes in
the flow rate.
Avoid abrupt pressure buildup in the system. This can lead to damage.
Charge the system with pressure only after establishing the electrical connections.
Be sure to have an inert gas (e.g. nitrogen) available for flushing in the case of dangerous, abrasive
or corrosive gases.