Specifications

RED SCARLET-X OPERATION GUIDE
COPYRIGHT © 2013 RED.COM, INC
955-0003_v4.0, Rev-A | 145
To calibrate the sensor the image needs to be as dark as possible:
At a minimum put a lens cap on the lens (WARNING: Not all lens-caps completely block IR. Lens caps
are made to protect the lens, not necessarily to block light.)
Preferably, take off the lens, and use the lens-mount cap provided by RED.
If possible, use a dark room.
Just closing the Iris is NOT sufficient.
Additionally a SSD needs to be inserted in the camera to store the raw images during calibration.
Calibration is required:
If the temperature deviates significantly from the currently used calibration, it is recommended to
recalibrate the sensor.
If very long exposure (~longer then 1/24 Sec) is used, it is recommended to recalibrate.
After Calibration with a lens cap on and ISO set to maximum, the sensor should show a uniform noise profile
over the whole area without any falloffs on any side. You can check at which temperature/exposure time the
currently used calibration was taken on under SETTINGS > MAINTENANCE > SYSTEM STATUS under
Calibration.
Calibration maps will still be available after performing a system restore under SETTINGS > MAINTENANCE
> RESTORE SYSTEM; however the factory calibration map will be reselected.