SIM™ PATIENT ENVIRONMENT
The SIM™ system medical electrical system is comprised of a combination of medical electrical equipment,
information technology electrical equipment.
Simavita has tested, certified and classified the SIM™ pod and SIM™ Sensor as medical electrical equipment
and has tested, certified and classified the SIM™ dock as information technology equipment
WARNING
In order to ensure that patient safety is maintained at all times the SIM™ dock or other electrical
equipment must not be placed within the patient environment as defined and illustrated below.
In case of emergency, disconnect the equipment from power supply mains.
In order to ensure that patient safety is maintained at all times, the operator must not touch the
Resident at the same time if the patient is connected to the SIM™ pod and SIM™ sensor.