User's Manual

Information to the equipment
installer
All Renishaw equipment is designed to comply
with the relevant EC and FCC regulatory
requirements. It is the responsibility of the
equipment installer to ensure that the following
guidelines are adhered to, in order for the product
to function in accordance with these regulations:
Any interface must be installed in a
position away from any potential sources
of electrical noise, i.e. power
transformers, servo drives etc.
All ground connections should be
connected to the machine 'star point' (the
'star point' is a single point return for all
equipment ground and screen cables).
Failure to adhere to this can cause a
potential difference between grounds.
All screens must be connected as outlined
in the user instructions.
Cables must not be routed alongside high
current sources (i.e. motor power supply
cables etc), or be near high speed data
lines.
Cable lengths should always be kept to a
minimum.
Equipment operation
If the equipment is used in a way not specified by
the manufacturer, the protection provided by the
equipment may be impaired.