User's Manual

RLP40 installation guide
6.2
Fault finding
Symptom Cause Action
Machine stops
unexpectedly during a
probing cycle.
Radio link failure/RLP40 out of
range.
Check interface/receiver and
remove obstruction.
RMI receiver/machine fault. Refer to receiver/machine user’s
guide.
Dead batteries. Change batteries.
Probe unable to find target
surface.
Check that part is correctly
positioned and that stylus has not
broken.
Stylus not given sufficient time to
settle from a rapid deceleration.
Add a short dwell before the
probing move (length of dwell will
depend on stylus length and rate
of deceleration). Maximum dwell is
one second.
Probe crashes.
Workpiece obstructing probe path. Review probing software.
Probe length offset missing Review probing software.