User's Manual

22
Replacing the RMI cover
1. Before replacing the cover, check for any
damage to screws or scratch marks
which could prevent sealing.
2. Ensure that the 'O' ring seating on the RMI
body is clean, and there are no scratch
marks which could prevent complete
sealing.
3 Ensure that the cover, antenna contacts
and 'O' ring are clean.
4. Place cover complete with 'O' ring
onto the RMI body.
Note :
The 'O' ring should be lubricated with
silicone grease to prevent nicking.
Do not get any grease on the antenna
contacts.
5. Tighten each captive screw a few turns at
a time, to pull the cover down evenly.
Screw torque is 1,4 Nm (1.03 lbf.ft).
It is not necessary to remove the RMI from
the machine when adjusting the switches or
installing new parts. For torque settings see
Screw torque values.
Removing the RMI cover
1. Clean RMI to ensure no debris enters unit.
2. Unscrew the three cover screws evenly
(T10 torx key). Do not remove screws
from cover.
3. When removing cover, do not twist or
rotate by hand.
RMI cover
RMI cover