User's Manual

27
Fault-finding
Probe not switched on
Probe out of range
Wiring fault
Output over-current
Low RMP60 batteries
Local radio interference
RMI error LED illuminated
during intended probe
cycle
All RMI LEDs flashing
RMI low battery LED lit
Reduced range
Check configuration and alter
as required
Check position of RMI,
see Operating envelope
(RMP60 User’s guide)
Check wiring
Check wiring, turn power to
RMI off and on again to reset
Change RMP60 batteries soon
Identify and move
Symptom Cause Action