RMP60-RMI partnership
Note
To check that turn on and off settings have not
accidentally been changed, remove and insert
batteries to review the current probe settings.
Note
When holding the RMP60, do NOT wrap a
hand, or anything else, around the glass
window.
Note
When the RMP60 and RMI become partners
the RMI records the RMP60 identity. It is not
possible for an RMI to be partnered with more
than one RMP60 in ‘Standard Mode’ - see
Multiple Probe Mode.
It is possible for an RMP60 to be partnered
with more than one RMI, but the system will
not function correctly.
8. Release the RMP60 stylus and trigger it a
couple of times. This causes the RMP60 to
go into (and out of) acquisition mode.
9. The RMI signal LED will change to
repeatedly flashing red and yellow (for the
remainder of the 10 second interval),
indicating a successful partnering.
10. Leave the RMP60 for 20 seconds to go
into standby.
11. The system is ready to use.