User's Manual

40
Fault-finding
Poor measurement
results.
Symptom Cause Action
Debris on part or stylus.
Repeatability of probe
into spindle.
Loose probe to shank
mounting or stylus.
Offsets not being updated.
Calibrated feature has moved.
Measurement occurs as
stylus leaves surface.
Calibration and probing
speeds different.
Clean and recalibrate.
Verify by repeated toolchange
and single point move.
Check and tighten as
required, recalibrate.
Review software.
Check.
Review software.
Review software.