Technical Specs

17
Practice Mode and Training Mode Measurements
Measure
Range
Accuracy
Precision
Angle
0-180°
+/- 0. 2°
Start Angle
0-180°
+/- 0. 2°
Max Angle
0-180°
+/- 0.2°
Session Score
0-100
+/- 0.5
1
Final Score
0-100
+/- 0.5
1
Storing Your leva
Storage of the leva PDHS should not exceed the following conditions:
Temperature: +5°C to +40°C
Humidity: 90% RH non-condensing
Pressure/Altitude: 3000m or 70kPa
Operating Your leva
Operation of the leva PDHS should not exceed the following
conditions:
Temperature: +5°C to +40°C
Humidity: 90% RH non-condensing
Pressure: 700hPa to 1060hPa
Operating Range: <3.2 ft (1 meter)
Expected Service Life
The expected service life of the leva PDHS is 1 year.
The leva PDHS requires periodic battery changes to maintain
reliability (see Replacing Batteries in the Maintenance Section).
App updates will be pushed to users periodically as needed
through the App Storeor Google Play.
Maintenance
Use the following instructions to ensure proper care and maintenance of
your leva PDHS.
Cleaning the Probe:
To ensure hygienic safety, you should clean and dry your leva probe before
and after each treatment session.
Clean the leva probe with mild, warm, soapy water.