User Manual

22
Device without humidification and SlimLine air tubing / Device with humidification and SlimLine air tubing
Breath
rate
Inspiratory pressure (cm H
2
O [hPa]) (Means, Standard Deviations)
6
10
16
21
25
10 BPM
-0.26, 0.01 / -0.52,
0.01
-0.25, 0.02 / -0.53,
0.02
-0.24, 0.02 / -0.53,
0.01
-0.25, 0.02 / -0.54,
0.02
-0.20, 0.02 / -0.51,
0.02
15 BPM
-0.26, 0.01 / -0.51,
0.01
-0.25, 0.01 / -0.54,
0.01
-0.26, 0.01 / -0.56,
0.01
-0.31, 0.03 / -0.58,
0.02
-0.30, 0.05 / -0.60,
0.03
20 BPM
-0.25, 0.02 / -0.52,
0.01
-0.29, 0.02 / -0.58,
0.01
-0.34, 0.02 / -0.62,
0.01
-0.36, 0.02 / -0.67,
0.02
-0.36, 0.03 / -0.69,
0.02
Breath
rate
Expiratory pressure (cm H
2
O [hPa]) (Means, Standard Deviations)
2
6
12
17
21
10 BPM
-0.28, 0.01 / -0.43,
0.01
-0.30, 0.03 / -0.50,
0.01
-0.30, 0.01 / -0.54,
0.01
-0.33, 0.01 / -0.58,
0.01
-0.34, 0.01 / -0.60,
0.02
15 BPM
-0.24, 0.02 / -0.37,
0.01
-0.29, 0.02 / -0.47,
0.01
-0.35, 0.01 / -0.55,
0.01
-0.38, 0.01 / -0.62,
0.02
-0.42, 0.02 / -0.66,
0.01
20 BPM
0.05, 0.21 / -0.38,
0.01
-0.31, 0.02 / -0.50,
0.02
-0.37, 0.02 / -0.57,
0.02
-0.43, 0.02 / -0.65,
0.02
-0.48, 0.02 / -0.68,
0.02
Note: The table above is based on data that covers between 60.1 and 88.8% of the inspiratory phase and 66.1 and 93.4% of the expiratory
phase durations. These data time slots start immediately after the initial transient overshoot/undershoot periods and end at the point that
flow diminishes to an equivalent absolute value of its starting point, towards the end of the breath phases (this corresponds to the % ranges
of values given immediately above).
Flow (maximum) at set pressures
The following are measured accordingly to ISO 80601-2-70:2015 at the end of the specified air tubing:
Pressure
cm H
2
O (hPa)
AirCurve 10 and
Standard
L/min
AirCurve 10,
humidification and
Standard
L/min
AirCurve 10 and
SlimLine
L/min
AirCurve 10,
humidification and
ClimateLineAir
L/min
4
180
143
162
151
8
168
135
151
142
12
157
136
140
135
16
144
134
128
121
20
131
123
117
109
25
120
115
96
84