ENGLISH S8 Escape™ System USER’S MANUAL English 338215-Eng/3 05 08 1
338215r3 Illustrations.
Page 2 H © 2005 ResMed Ltd. HUMIDAIRE 3i and S8 ESCAPE / HUMIDAIRE 3i et S8 ESCAPE / HUMIDAIRE 3i y S8 ESCAPE / HUMIDAIRE 3i e S8 ESCAPE F 1 USING A HUMIDAIRE™ / UTILISATION D'UN HUMIDAIRE / USO DE UN HUMIDAIRE / UTILIZAÇÃO DE UM HUMIDAIRE 5 4 6 3 7 8 1 0 5 4 5 6 4 3 6 3 7 2 7 2 8 1 8 1 0 0 2 DATA CARD MODULE / MODULE DE CARTE DE DONNÉES / MÓDULO TARJETA DE DATOS / COMO USAR O CARTÃO DE DADOS 1 3 2 Note: Some products not available in all regions.
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 YOUR RESPONSIBILITY 5 MEDICAL INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 ABOUT THE S8 ESCAPE™ CONTRAINDICATIONS WARNINGS CAUTIONS ADVERSE EFFECTS ENGLISH CONTENTS 5 6 6 7 8 THE S8 ESCAPE SYSTEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SYSTEM SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 S8 ESCAPE ACCESSORIES 29 29 LIMITED WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 INDEX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Thank you for choosing the S8 ESCAPE™ flow generator. The S8 ESCAPE is a compact, all-in-one package with an inbuilt power supply and the option of integrated data management and humidification. This user manual contains the information you need for the correct use of your S8 ESCAPE.
CONTRAINDICATIONS The S8 ESCAPE is not a life support device and may stop operating during power failure or certain fault conditions. It should not be used by patients who are dependent on continuous therapy. Positive airway pressure therapy may be contraindicated in some patients with the following pre-existing conditions: • severe bullous lung disease • pneumothorax • pathologically low blood pressure • dehydration • cerebrospinal fluid leak, recent cranial surgery, or trauma.
If oxygen is used with this device, the oxygen flow must be turned off when the device is not operating. If oxygen has been left on, turn off the flow generator, then wait 30 minutes before turning on the flow generatory again. Explanation: When the CPAP device is not in operation and the oxygen flow is left on, oxygen delivered into the air delivery tubing may accumulate within the CPAP machine enclosure and create a risk of fire. This applies to most types of CPAP machines.
ADVERSE EFFECTS Patients should report unusual chest pain, severe headache, or increased breathlessness to their prescribing physician. An acute upper respiratory tract infection may require temporary discontinuation of treatment. The following side effects may arise during the course of therapy with the CPAP device: • drying of the nose, mouth, or throat • nosebleed • bloating • ear or sinus discomfort • eye irritation • skin rashes • chest discomfort.
The S8 ESCAPE device is part of a system that consists of five elements: • S8 ESCAPE flow generator • Mask • Humidifier • Data management products • Accessories. ENGLISH THE S8 ESCAPE SYSTEM S8 ESCAPE Please refer to the illustrations in section A of the illustration sheet. FRONT VIEW (A-1) Shows the keypad and LCD screen, and the air outlet. REAR VIEW (A-2) Shows where the module attaches, and the location of the air filter and power sockets.
FULL FACE MASKS • • • ! MIRAGE™ FULL FACE MASK MIRAGE™ FULL FACE MASK SERIES 2 ULTRA MIRAGE™ FULL FACE MASK. WARNING Only ResMed mask systems are compatible for use with the S8 Escape. Please refer to “Warnings” on page 6. HUMIDIFIERS A humidifier may be required if you are experiencing dryness of the nose, throat, or mouth.
Please refer to the illustrations in section D of the illustration sheet. SET UP THE S8 ESCAPE 1 ENGLISH HOW TO USE THE S8 ESCAPE Place the S8 ESCAPE on a table near the head of your bed. ! CAUTION Be careful not to place the device where it can be bumped or where someone is likely to trip over the power cord. Note: You can also place your S8 Escape on the floor beside or underneath your bed.
2 Lie down and arrange the air tubing so that it is free to move if you turn in your sleep. ! CAUTION Do not leave long lengths of air tubing around the top of your bed. They could twist around your head or neck while you are sleeping. • Make sure the area around the flow generator is dry and clean. It should also be clear of bedding, clothes, and other potential blockages. • 3 Alter the ramp time if required (see “How to Change the Ramp Time” on page 17).
ATTACH A HUMIDAIRE 3i HUMIDIFIER Please refer to the illustration in section E of the illustration sheet, and your HumidAire 3i manual. The HUMIDAIRE 3i attaches to the front of the S8 ESCAPE to provide heated humidification. No other accessories are required for its use. ENGLISH HOW TO ATTACH A HUMIDIFIER Notes: a. Make sure the water chamber of the HumidAire 3i is empty before you attach or detach the humidifier. b.
ATTACH A RESMED PASSOVER HUMIDIFIER Please refer to the illustrations in section J of the illustration sheet. To connect your S8 ESCAPE to a ResMed PASSOVER humidifier you need medium sized 21” (52 cm) air tubing. 1 Fill the PASSOVER with water and connect the medium 21” (52 cm) and long 6’6” (2 m) air tubing (J-1). 2 Place the S8 ESCAPE on top of the PASSOVER (J-2) and connect the free end of the medium air tubing to the air outlet (J-3).
The control panel of the S8 ESCAPE includes an LCD screen and keypad for navigating through the menus and delivering treatment. ENGLISH HOW TO USE THE KEYPAD LCD screen Up key Right key Left key Down key Start/Stop key LCD SCREEN The LCD screen displays the menus and treatment screens. KEYPAD KEYS The S8 ESCAPE keypad has the following keys: Key Function Start/Stop • Starts or stops treatment. Up • Allows you to increase settings options and scroll through the S8 ESCAPE menu.
on when you press a key. Once the Start/Stop key has been pressed to start treatment, the LCD backlight stays on for two minutes. The keypad backlight is always on while the S8 ESCAPE is on. HOW TO CHANGE THE SETTINGS ON YOUR S8 ESCAPE You can change settings (such as Ramp Time, Altitude) on the S8 ESCAPE by using the keypad and LCD. When the S8 ESCAPE is in standby mode you can display a series of screens (a menu) on the LCD screen in order to view and change the settings for a particular function.
ENGLISH HOW TO EXIT OUT OF A MENU Press the Right key (Exit). This will take you back to the Patient Menu screen. HOW TO EXIT OUT OF AN ITEM WITHIN THE MENU Press the Right key (Exit). HOW TO CHANGE THE RAMP TIME During ramp time, the pressure increases from a low pressure to the prescribed treatment pressure. You can set the ramp time in 5-minute intervals, from 0 minutes (OFF) up to a maximum ramp time set by your clinician. The standby screen displays the current ramp time in the right-hand corner.
MESSAGES ON THE S8 ESCAPE LCD Your clinician may have set your S8 ESCAPE to remind you about important events, such as when to replace your mask, when to insert your Data Card (if your device is Data Card enabled) and so on. The reminder message is displayed on the LCD and is visible if the device is not delivering therapy. The backlight on the LCD flashes when a message is displayed.
Note: Do not remove modules while power is connected to the S8 Escape. Your S8 ESCAPE will have one of two modules fitted to the back: a blank module or a RESSCAN™ Data Card module. ENGLISH S8 MODULES USING A BLANK MODULE Please refer to the illustrations in section F of the illustration sheet. The blank module is a protective cover on the S8 ESCAPE. At times, you may wish to remove or attach this module, for example to use the serial adapter.
1 INSERT THE DATA CARD • Switch on the S8 ESCAPE and wait until you see the standby (Ramp) screen. • Hold the Data Card with the arrow facing up and insert it into the slot in the Data Card module (H-1). • Push the card in smoothly until it stops moving (H-2) and messages start appearing on the LCD. 2 COPY DATA ONTO THE DATA CARD Data copying starts automatically when the Data Card is inserted into the Data Card module.
WARNING If your clinician has told you to use this Data Card to update the settings on your device and the “Settings Success” message does not appear, contact your clinician immediately.
CLEANING AND MAINTENANCE You should regularly carry out the cleaning and maintenance described in this section. Refer to your mask manual for detailed instructions. DAILY CLEANING Mask Clean the mask according to the instructions supplied with the mask. Air tubing Disconnect the air tubing from the S8 ESCAPE (and humidifier, if used) and hang the tubing and mask in a clean, dry place until next use.
To replace the air filter: 1. Remove the air filter cover at the back of the S8 ESCAPE . ENGLISH REPLACING THE AIR FILTER Air filter cover 2. Remove and discard the old air filter. 3. Insert a new filter with the blue tinted side facing out from the device. 4. Replace the air filter cover. Inspect the air filter every month to check if it is blocked by dirt or contains holes.
FREQUENTLY ASKED QUESTIONS HOW WILL IT FEEL WHEN I AM BREATHING WITH THE S8 ESCAPE? When you first use your S8 ESCAPE, you may feel uncomfortable breathing against the airflow. This is normal. You will continue to breathe normally while you sleep. Take time to adjust to this new sensation. WHAT DO I DO WHEN I WANT TO GET UP IN THE NIGHT? If you need to get up during the night, remove your mask and stop therapy. Always remember to restart therapy and put your mask on when you return to bed.
If you are moving or traveling to a higher or lower altitude, see “How to change the Altitude setting” on page 17. On an aircraft Please consult the medical services department of your carrier if you intend to use your S8 ESCAPE on an aircraft. The S8 ESCAPE complies with US FCC Part 15, Class B requirements. ENGLISH Altitude changes Note: Do not use your S8 Escape while the aircraft is taking off or landing.
TROUBLESHOOTING If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the device. Problem Possible Cause Solution No display. Power not connected. Ensure the appropriate power cord is connected to your S8 Escape and the power switch (if available) is on. Insufficient air delivered from S8 Escape. Ramp Time is in use. Wait for air pressure to build up or change ramp time. Air filter is dirty.
Possible Cause Solution Display error message: Exxxx Call Service (where xxxx defines an error) or MOTOR FAULT Call Service or UPLOAD LANGUAGES Call Service or DATA LOST Call Service Component failure. Call Service. The following message is displayed on the LCD after you try to update settings using the Data Card: Card Error Remove Card The Data Card is not inserted correctly. Ensure that the Data Card is inserted with the arrow facing up and pointing towards the device.
28 Problem Possible Cause Solution The following message is displayed on the LCD after you try to update the settings using the Data Card: Settings Error Remove Card There is a data error on the Data Card. Contact your clinician immediately. The following message is NOT displayed on the LCD after you try to update the settings using the Data Card: Settings Success Remove Card The settings were not updated. Contact your clinician immediately.
S8 ESCAPE Performance: Operating pressure range: 4 to 20 cm H2O Dimensions (H x W x D): 4.6” x 6.5” x 5.7” (112 mm x 164 mm x 145 mm) Weight: 2.9lb (1.3 kg) ENGLISH SYSTEM SPECIFICATIONS Power Supply Input range for S8 ESCAPE with HUMIDAIRE 3i:100–240V, 50/60Hz; 110V, 400Hz; 2.5A < 140VA (110W) (maximum power consumption). Instantaneous peak power consumption <340 VA. • Refer to the DC-12 converter instructions for DC ratings. Only use the ResMed DC-12 converter for DC input.
DATA CARD MODULE Dimensions (H x W x D): 1.95” x 1.66” x 0.47” (49.7 mm x 42.
Guidance and manufacturer’s declaration—electromagnetic emissions The S8 series of devices is intended for use in the electromagnetic environment specified below. The customer or the user of the S8 device should assure that the device is used in such an environment.
Guidance and manufacturer’s declaration—electromagnetic immunity The S8 series of devices is intended for use in the electromagnetic environment specified below. The customer or the user of the S8 device should assure that the device is used in such an environment.
The S8 series of devices is intended for use in the electromagnetic environment specified below. The customer or the user of the S8 device should assure that the device is used in such an environment.
Recommended separation distances between portable and mobile RF communications equipment and the S8 series of devices The S8 series of devices is intended for use in an environment in which radiated RF disturbances are controlled.
ResMed warrants that your ResMed product shall be free from defects in material and workmanship for the period specified below from the date of purchase by the initial consumer. This warranty is not transferable. Product Warranty Period ResMed humidifiers, ResControl™, ResLink™ 1 Year ResMed flow generators 2 Years Accessories, mask systems (including mask frame, cushion, headgear and tubing). Excludes single-use devices.
36
A accessories 10 air filter 9, 26 replacing 23 air pressure 26 air tubing 9 cleaning 22 connect to air outlet 11 kinked 26 positioning 12 aircraft 25 airflow breathing with 24 altitude setting 17, 25, 26 B backlight LCD and keypad 15 C carry bag 9 cleaning periodic 22 contraindications 6 CPAP 5 D daily cleaning 22 Data Card 19 copying data 20 inserting 20 mailback envelope 20 removing 20 data card 27 Data Card module 19 DC-12 converter 25 down key 15 dryness mouth, nose or throat 24 E error message 27 F fre
N nasal irritation 24 P Passover attaching 14 power cord 9, 26 connecting 11 power sockets 9 power sources 25 power supply 29 R ramp time 12, 17, 26 reminder replace mask 18 reminders 18 ResScan Data Card module 10 right key 15 S S8 Escape dimensions 29, 30 servicing 23 settings 16 set-up 11 start treatment 12 stop treatment 12 travelling with 24 setting up 11 software version number 17 standby screen 11 start/stop key 15 SW 17 system specifications 29 T troubleshooting 26 U up key 15 Usage 17 use on aircra