Operator`s manual

16
Monitoring
Verify that the pulse oximeter sensor is properly positioned on the patient.
Ensure that the pulse oximeter is sensing adequate pulse quality by:
verifying that the Pulse Quality display is blinking green and
verifying that the Pulse Rate and SpO
2
displays are displaying
readings and
verifying that blinking of the Pulse Quality display is in time with the
pulse rate for at least 10 seconds
If the Pulse Quality display is blinking red or amber or is blinking
erratically, reposition the sensor or replace the sensor.
If the sensor is not properly positioned, or no sensor is attached to the
pulse oximeter after startup (a few seconds after powering on), both the
SpO
2
and Pulse Rate displays will display a single dash until a valid pulse
signal is detected.