Operator`s manual

20
Visual Indicators
The following table describes visual indicators and conditions.
Condition Visible Indication
Pulse Waveform Signal
is inadequate
Pulse Quality LED blinks red
•SpO
2
and heart rate LEDs display
dashes after 10 seconds
Sensor fault (i.e. sensor
disconnect, failure, or
incompatibility with the
monitor)
Pulse Quality LED blinks
Dash (-) appears in the leftmost position
of the SpO
2
display
•SpO
2
and Heart Rate numeric displays
freeze for 10 seconds
Inadequate Signal
(Sensor misalignment,
ambient light, nail
polish/artificial nails,
etc.)
Pulse Quality LED is blank
Dash (-) appears in the leftmost position
of the SpO
2
display
•SpO
2
and Heart Rate numeric displays
freeze for 10 seconds
Inadequate SpO
2
or
pulse rate data
(excessive motion or
erratic heart rate, etc.)
more than 20 seconds
Dash (-) appears in SpO
2
and pulse rate
displays
Pulse rate data not
updated for more than
30 seconds
Pulse rate numeric display becomes
dashes
Low Battery
Low Battery indicator is solid amber
LED. No other displays are affected.
Critically Low Battery
Critically Low Battery indicator is
flashing
•SpO
2
and pulse rate LEDs are blank
Pulse Quality display is solid red or