Table 4: Recommended Separation Distances
This table details the recommended separation distances between portable and
mobile RF communications equipment and this device.
This device is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. Users of this device can help prevent
electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communication equipment (transmitters) and the device
as recommended below, according to maximum output power of the
communications equipment.
Separation Distance According to
Frequency of Transmitter
Rated Maximum
Output Power of
Transmitter W
150 kHz to
80 MHz
80 MHz to
800 MHz
800 MHz to 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.2 1.2 2.3
10 3.7 3.7 7.4
100 12 12 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTES:
At 80 MHz and 800MHz, the separation distance for the higher frequency
range applies.
These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects, and people.
d 1.17 P=
d 1.17 P=
d 2.33 P=