Operator`s manual

41
Indicators
Pulse Quality Display: LED, tricolor
Numeric Displays: 3-digit 7-segment LEDs, red
Low Battery Indicator: LED, amber
Temperature (Operating) -20 to +50°C (-4 to +122°F)
Temperature (Storage/Transportation): -30 to +50°C (-22 to +122°F)
Humidity (Operating) 10 to 90% noncondensing
Humidity (Storage/Transportation): 10 to 95% noncondensing
Altitude (Operating) Up to 12,000 meters (40,000 feet)
Altitude (Hyperbaric Pressure): Up to 4 atmospheres
Power Requirements Four 1.5V AA-size alkaline batteries
(80 hours typical operation) or NiMH
rechargeable battery pack (40 hours
typical operation)
Dimensions 13.8 cm H x 7.0 cm W x 3.2 cm D
(5.4 in H x 2.8 in W x 1.3 in D)
Weight 210 g (7.4 oz) (with alkaline batteries)
230 g (8.1 oz) (with NiMH
rechargeable battery pack)
Classifications per IEC 60601-1 / CSA601.1 / UL60601-1 30EM
Type of Protection: Internally powered (on battery power)
Degree of Protection: Type BF-Applied Part
Mode of Operation: Continuous
Enclosure Degree of Ingress
Protection
IPX2