7.3 Cl e a n I n g T h e TU b I n g
Cleanthetubingdaily.Disconnecttheexibletubingfromthedevice.Gentlywashthetubingin
a solution of warm water and a mild detergent. Rinse thoroughly. Air dry.
7.4 Cl e a n I n g T h e Wa T e r Ch a M b e r
RefertoFigure7-1whencleaningthewaterchamber.
Protective Cover
(Hand wash only!)
Flapper (Soft) Valve
(Hand wash only!)
Water Chamber with
Removable Base Plate
fI g U r e 7–1 Cl e a n I n g T h e Wa T e r Ch a M b e r
NOTE: eFlapper(Soft)Valvehasaoneyearlifespan.Youmustreplacethisvalveafterone
year.
NOTE: Handwashingcanbeperformeddaily.Dishwashingcanbeperformedonceaweek.
WARNING: Empty and clean the water chamber daily to prevent mold and bacteria growth.
WARNING: Allow the water in the chamber to cool to room temperature before removing the
chamber from the humidifier.
1. Turnthetherapydeviceoandallowapproximately15minutesfortheheaterplateand
water to cool.
2. Disconnectthetubingfromthewaterchamber.
3. Slide the water chamber out of the humidifier platform. Empty any remaining water.
4. Gentlyremovethebaseofthechamberwithyourhands,beingcarefulnottodamagethe
rubber seal.
5. Removetheappervalvefromtheairinletport;uncliptheprotectivecoverfromthetopof
the water chamber. Wash these two parts by hand only in a solution of warm water and mild
dishwashing soap. Rinse with clean water and allow them to air dry.