ap p e n d I x a: eMC In f o r M a T I o n
gU I d a n C e a n d Ma n U f a C T U r e r s de C l a r a T I o n - el e C T r o M a g n e T I C eM I s s I o n s is device is intended for use in the
electromagnetic environment specified below. e user of this device should make sure it is used in such an environment.
eM I s s I o n s Te s T Co M p l I a n C e el e C T r o M a g n e T I C en v I r o n M e n T - gU I d a n C e
RF emissions
CISPR 11
Group 1 The device uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuations/Flicker emissions
IEC 61000-3-3
Complies
gU I d a n C e a n d Ma n U f a C T U r e r s de C l a r a T I o n - el e C T r o M a g n e T I C IM M U n I T y is device is intended for use in the
electromagnetic environment specified below. e user of this device should make sure it is used in such an environment.
IM M U n I T y Te s T IeC 60601 Te s T
le v e l
Co M p l I a n C e le v e l el e C T r o M a g n e T I C en v I r o n M e n T -
gU I d a n C e
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic
tile. If oors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
Transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input-output lines
±2 kV for supply mains
±1 kV for input/output lines
Mains power quality should be that of a
typical home or hospital environment.
Surge
IEC 61000-4-5
±1 kV dierential mode
±2 kV common mode
±1 kV dierential mode
±2 kV for common mode
Mains power quality should be that of a
typical home or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for
0.5 cycle
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in U
T
) for
5 sec
<5% U
T
(>95% dip in U
T
) for
0.5 cycle
40% U
T
(60% dip in U
T
) for 5 cycles
70% U
T
(30% dip in U
T
) for 25
cycles
<5% U
T
(>95% dip in U
T
) for
5 sec
Mains power quality should be that of a
typical home or hospital environment. If
the user of the device requires continued
operation during power mains interruptions,
it is recommended that the device be
powered from an uninterruptible power
supply or a battery.
Power frequency (50/60 Hz)
magnetic eld
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic elds should be
at levels characteristic of a typical location in
a typical hospital or home environment.
NOTE: U