User's Manual

RXiBreeze PAP System User Manual
17
Figure3 Patient Comfort Setting Interface
Icon
Text
Description
Preheat
Turn Preheat function on or off. When the Preheat is on, the
humidifier starts preheating in the standby mode and the
maximum preheating time is 30 minutes. In the therapy mode,
the preheating stops.
Option: ON / OFF Default: OFF
Note:
1. If the humidifier water level below the limit, Preheat function
will be turned off automatically.
2. If humidifier function is turned off, Preheat function will be
disable.
3. Preheat function is not available in the RXiBreeze 20C.
P Ramp
Set the starting pressure of Ramp.
Setting Range: 3 cmH
2
O-Setting pressure, 0.5 cmH
2
O
increments.
Default: 4 cmH
2
O
Ramp
Time
Set the increase time from Ramp pressure to the setting
therapy pressure.
Option: 0-60 mins, 5 mins increments
Default: 15 mins
Note: if Ramp Time sets to 0 min, Ramp function will be off.
Auto Start
Turn Auto Start function on or off.
When the Auto Start function is on, the system will start
therapy automatically if a breath with mask is detected.
Option: On / OFF Default: ON
Auto Stop
Turn Auto Stop function on or off.
When the Auto Stop function is on, if the mask is removed
more than 5 seconds, the therapy mode will stop.
Option: On / OFF Default: ON