Alert
Type
Possible Cause
Action
Software update
underway, Do not
cut off the power!
Notification
Software update.
No action.
System error code:
XXXX Please try to
restart, please
contact the supplier
if repeat.
Alert 2
1. Pressure sensor
failure, flow sensor
failure, blower failure in
therapy state; 2. Power
board short-circuit 3.
Humidifier heating
circuit short-circuit
Please try to restart,
please contact the
supplier if repeat.
The respiration
tube expired,
please replace.
Notification
In standby state, the
usage time exceed the
respiration tube setting
service life.
Click “Confirm”,
replace the respiration
tube.
The water tank
expired, please
replace.
Notification
In standby state, the
usage time exceed the
water tank setting
service life.
Click “Confirm”,
replace the water
tank.
The filter expired,
please replace.
Notification
In standby state, the
usage time exceed the
filter setting service
life.
Click “Confirm”,
replace the filter
The mask expired,
please replace.
Notification
In standby state, the
usage time exceed the
mask setting service
life.
Click “Confirm”,
replace the mask
It is time for device
maintenance,
Please contact the
service provider for
device
maintenance.
Notification
In standby state, the
usage time exceed the
setting device
maintenance time.
Click “Confirm”,
contact the service
provider for device
maintenance.
Apnea.
Alert 1
An apnea is detected
and exceeds the
setting duration.
Check the respiration
tube or adjust therapy
parameter setting.
Mask vent holes
blocked.
Alert 1
Mask vent holes is
blocked.
Check the mask.
System Drying
Notification
The device generates
a small flow to dry the
system when it is the
standby mode after 30
No action.